65
Measurement of salinity
Measurement range / Resolution
0.000…1.999g/l / 1mg/l
2.00…19.99g/l / 10mg/l
20.0…199.9 g/l / 0.1 g/l
Accuracy (salinity)
±
0.5%
±
1digit
Automatic/manual temperature compensation
0...100°C with
α
T
= 0.00…4.00%/°C
Reference temperature
0…50°C (
default 20°C or 25°C
)
χ
/TDS conversion factor
0.4…0.8
Preset cell constant K (cm
-1
)
0.01 - 0.1 - 0.5 - 0.7 - 1.0 - 10.0
Cell constant K (cm
-1
) that can be set by the user
0.01…20.00
Standard solutions automatically detected (@25°C)
147
μ
S/cm
1413
μ
S/cm
12880
μ
S/cm
111800
μ
S/cm
•
Measurement of temperature by Instrument
Pt100 measurement range
-50…+150°C
Pt1000 measurement range
-50…+150°C
Resolution 0.1°C
Accuracy
±
0.1°C
±
1digit
Drift after 1 year
0.1°C/year
Summary of Contents for HD2205.2