38
Sensor
Card connection
Jumper to close
Pt1000 2 wires
JP4
4
JP3
3
JP1
1
JP2
2
Pt1000
2 wires
JP2
Ensure the welds are clean and perfect. Once the welding operation is complete, close the two
shells, insert the gasket in the module, and screw the fairlead. At the other end of the module, enter
the ring with the O-ring as indicated in the picture.
Make sure the cable is not twisted while you are screwing the fairlead. Now the probe is ready.
Direct connection of the 4 wire Pt100 sensor to a DIN45326 connector
The
4 wire Pt100
sensor can be
welded directly on the DIN45326 fe-
male connector's pins, without using
the TP47 card. The 4 wires of the
Pt100 are welded as outlined in the
scheme on the side.
The Pt100 probe is detected during
turn on: insert the probe when the in-
strument is off, and switch it on.
This connection does not guarantee
the IP66 protection degree.
Sensor
Direct soldering to the connector
Pt100 4 wires
View of the soldering side
of the flying female connector