![Delta OHM HD2102.1 Manual Download Page 10](http://html1.mh-extra.com/html/delta-ohm/hd2102-1/hd2102-1_manual_2478293010.webp)
before entering the menu. By pressing the UNIT key, it is possible to select the coefficient 3,
6 and 9 that multiplies the displayed value respectively by 10
3
, 10
6
, 10
9
.
2.
TIME LIMT HOUR
: indicates the number of hours for the integration time limit, after
which the integral calculation is stopped.
3.
TIME LIMT MIN
: indicates the number of minutes for the integration time limit, after
which the integral calculation is stopped.
4.
TIME LIMT SEC
: indicates the number of seconds for the integration time limit, after
which the integral calculation is stopped.
5.
Management of memorized data (only HD2102.2)
: the message
“>>>_LOG_DUMP_or_ERAS
” (Transfer data or erase) is scrolled in the comment line.
The center figure reports the number of free memory pages (FREE). All memory data are
permanently erased by pressing SERIAL/EraseLOG. By pressing LOG/DumpLOG, the data
transfer of the logged data on the serial port is started: the “BAUD-RATE” must have
previously been set to the maximum value (please see the menu items described below and
the paragraph "STORING AND TRANSFERRING DATA TO A PERSONAL
COMPUTER" on page 21).
6.
Sleep_Mode_LOG (Automatic turning off during recording) (only HD2102.2):
this
function controls the instrument's automatic turning off during logging, occurring between
the capture of a sample and the next one. When the interval is lower than 60 seconds, the
instrument will always remain on. With intervals greater than or equal to 60 seconds, it is
possible to turn off the instrument between loggings: it will turn on at the moment of
sampling and will turn off immediately afterwards, thus increasing the battery life. Using the
arrows select
YES
and confirm using
ENTER
in order to enable the automatic turning off,
select
NO
and confirm to disable it and keep the instrument on continuously.
Note:
even if
Sleep_Mode_LOG=YES
is selected, the instrument does not turn off for less
than one minute intervals.
7.
Print and log interval
: sets the interval in seconds between two loggings or data transfers
to the serial port. The interval can be set from 0 to 3600 seconds (1 hour).
If the value 0 is
set, SERIAL works on command: the sending of data to the serial port is performed
each time the key is pressed
. Recording (LOG) is performed with one second intervals
even if the interval is set to 0. With an interval from 1 to 3600s, continuous data transfer is
started when the SERIAL key is pressed. To end the recording (LOG) and
continuous
data
transfer operations (SERIAL with an interval greater than 0), press the same key again.
8.
YEAR
: to set the current year. Use the arrows to modify this parameter and confirm using
ENTER.
9.
MNTH (month)
: to set the current month. Use the arrows to modify this parameter and
confirm using ENTER.
10.
DAY
: to set the current day. Use the arrows to modify this parameter and confirm using
ENTER.
11.
HOUR
: to set the current hour. Use the arrows to modify this parameter and confirm using
ENTER.
12.
MIN (minutes)
: to set the current minutes. In order to correctly synchronize the minute, it is
possible to reset the seconds by pressing the UNIT key. Use the arrows to set the current
minute plus one, and as soon as that minute is reached press UNIT: this synchronizes the
time to the second. Press ENTER to go onto the next item.
- -