T
HE
FUNCTION
The PRINT function sends the measurements taken in real time by the instrument inputs directly to
a PC. Print data units of measurements are the same as those used on the display. The function is
started by pressing SERIAL. The time interval between two consecutive prints can be set from 1
second to 1 hour (please see the
Print and log interval
menu item on page 9). If the print interval is
equal to 0, by pressing SERIAL the single data is sent to the connected device. If the print interval
is higher than 0, the data transfer continues until the operator stops it by pressing SERIAL again.
NOTES:
•
The print out is formatted across 24 columns.
•
Some keys are disabled during serial transmission. The following keys work: ON/OFF, HOLD,
FUNC (Max-Min-Avg) and LOG.
•
Pressing the HOLD, REL and FUNC keys has no effect on the printed data if these keys are
pressed
after
starting the printing, otherwise the following is valid.
•
The transfer started with the display in HOLD mode proceeds normally with the actual
measured values (that is, not in “HOLD” mode). Only the display is frozen to the values present
when the HOLD key was pressed.
•
The same is true for the Max-Min-Avg function.
•
If the serial transfer is started when the display is in REL mode, the relative values are
transferred.
•
It is possible to activate both the logging (LOG) and direct transfer (PRINT) functions at the
same time.
- -