WARNINGS
AND
OPERATING
INSTRUCTIONS
1.
Do not bend the probe connectors or force them upward or downward.
2.
Do not bend or force the contacts when inserting the probe connector into the instrument.
3.
The sensors and filters should not exceed the temperature limits established with consequent
irreparable degradation of their characteristics.
4.
Do not drop the probes: as this could cause irreparable damage.
5.
Avoid taking measurements in presence of high frequency sources, microwave ovens or large
magnetic fields; results may not be very reliable.
6.
The instrument is water resistant and IP67, but is not watertight and therefore should not be
immersed in water without closing the free connectors using caps. The probe connectors must
be fitted with sealing gaskets. Should the instrument fall into the water, check for any water
infiltration. Gently handle the instrument in such a way as to prevent any water infiltration from
the connectors' side.
INSTRUMENT
SIGNALS
AND
FAULTS
The following table lists all error indications and information displayed by the instrument and
supplied to the user in different operating situations:
Display indications
Explanation
ERR
This appears if the probe has already been detected by the instrument,
but is disconnected. At the same time an intermittent beep is issued.
PROB COMM LOST
This appears if the probe has already been detected by the instrument,
but is disconnected. At the same time an intermittent beep is issued.
OVER
Measurement overflow: indicates that the probe is measuring a value
exceeding the measuring range.
LOG
MEM
FULL
Memory full; the instrument cannot store further data, the memory
space is exhausted.
PROB
ERR
A probe with SICRAM module has been inserted when not admissible
for that specific instrument.
CAL
LOST
Program error: it appears after turning on for a few seconds. Contact
the instrument's supplier.
SYS
ERR
#
Instrument management program error. Contact the instrument's
supplier and communicate the numeric code # reported by the display.
BATT TOO LOW
CHNG NOW
Indication of insufficient battery charge appearing on turning on. The
instrument issues a long beep and turns off. Replace the batteries.
- -