![Delta OHM HD2102.1 Manual Download Page 31](http://html1.mh-extra.com/html/delta-ohm/hd2102-1/hd2102-1_manual_2478293031.webp)
IRRADIANCE measurement probe LP 471 UVB complete with SICRAM module and
equipped with the instrument
Measurement range (W/m
2
):
0.1
⋅
10
-3
… 999.9
⋅
10
-3
1.000…19.999 20.00…199.99
200.0…1999.9
Resolution (W/m
2
):
0.1
⋅
10
-3
0.001 0.01 0.1
Spectral range:
280nm…315nm (Peak 305nm)
Calibration uncertainty:
<5%
f'
1
(response according to the cosine
law):
<6%
f
3
(linearity):
<2%
f
4
(instrument reading error):
±
1digit
f
5
(fatigue):
<0.5%
Drift after 1 year:
<2%
Working temperature:
0…50
°
C
Typical response curve
0
0,1
0,2
0,3
0,4
0,5
0,6
0,7
0,8
0,9
1
250
260
270
280
290
300
310
320
330
340
350
λ
(nm)
R
e
la
tiv
e s
pe
ct
ral
r
es
pon
se
- -
31
Summary of Contents for HD2102.1