Delta OHM HD2102.1 Manual Download Page 21

STORING

 

AND

 

TRANSFERRING

 

DATA

 

TO

 

A

 

PERSONAL

 

COMPUTER 

The HD2102.1 and HD2102.2 instruments can be connected to a personal computer via an RS232C 
serial port, and exchange data and information through the DeltaLog9 software running in a 
Windows operating environment. The HD2102.2 can also use the USB connection. Both models 
can send in real time input measured values directly to a PC, through the PRINT function; the 
HD2102.2 can also store the values measured by using the 

Logging

 function (LOG key) in its 

internal memory. If necessary, the data stored in the memory can be transferred to a PC later. 

T

HE 

L

OGGING 

FUNCTION 

-

 ONLY FOR 

HD2102.2 

The 

Logging

 function allows the recording up to 38,000 measurements registered by the probe 

connected to the input. The time interval between two consecutive measurements can be set from 1 
second to 1 hour. The logging starts by pressing the LOG key and ends by pressing the same key 
again: the data memorized in this way form a continuous block of data.  
See the description of the menu item on page 9
If the automatic turning off option between two recordings (menu >> 

Sleep_Mode_LOG) 

is 

enabled, upon pressing the LOG key the instrument logs the first data and turns off. 15 seconds 
before the next logging instant, it turns on again to capture the new sample, and then turns off.  
The data stored in the memory can be transferred to a PC using the DUMP LOG command: press 
the MENU key until the “>>>_LOG_DUMP_or_ERAS” item is displayed and then press on LOG 
key. During data transfer the display shows the message DUMP; to stop the data transfer press ESC 
on the instrument or on the PC. 
 

C

LEARING THE MEMORY 

-

 ONLY FOR 

HD2102.2 

To clear memory use the Erase 

Log function (the MENU key until the 

“>>>_LOG_DUMP_or_ERAS” item is displayed and then the SERIAL/EraseLOG key). 
The instrument starts clearing the internal memory; at the end of the operation, it goes back to 
normal display. 
 
NOTES: 

 

Data transfer does not cause the memory to be erased; the operation can be repeated as many 
times as required.  

 

The logged data remain in memory apart from battery charge conditions. 

 

In order to print the data to a parallel interface printer, you must use a parallel-serial adaptor (not 
supplied). 

 

The direct connection between instrument and printer via a USB connector does not work. 

 

Some keys are disabled during logging. The following keys work: HOLD, FUNC (Max-Min-
Avg) and SERIAL. 

 

Pressing the HOLD, REL and FUNC keys has no effect on the logged data if these keys are 
pressed 

after 

starting the recording, otherwise the following is valid.  

 

The recording started with the display in HOLD mode proceeds normally with the actual 
measured values (that is, not in “HOLD” mode). Only the display is frozen to the values present 
when the HOLD key was pressed.  

 

The same is true for the Max-Min-Avg function. 

 

If the logging is started when the display is in REL mode, the relative values are logged. 

 

It is possible to activate both the logging (LOG) and direct transfer (PRINT) functions at the 
same time. 

-      - 

21

Summary of Contents for HD2102.1

Page 1: ...This situation can cause possible differences comparing this manual with the instrument you bought We regret in advance for any possible mistake in this manual Data drawings and descriptions included...

Page 2: ...Photo Radiometer HD2102 1 2...

Page 3: ...en the key is pressed in the menu decreases the current value 9 SERIAL key starts and ends data transfer to the serial communication port 10 MENU CLR Q t key allows access to and exit from the menu If...

Page 4: ...Photo Radiometer HD2102 2 4...

Page 5: ...ey during normal operation starts and ends the saving of the measured or integrated data Q T in the internal memory in the menu starts the data transfer from the instrument s memory to the PC 11 MENU...

Page 6: ...n the menu When exceeded these thresholds cause the instrument to stop the integral calculation The HD2102 2 instrument is a datalogger It memorizes up to 38 000 samples which can be transferred from...

Page 7: ...detection of the probe connected to the input and sets the instrument ready for normal measurement MIN cm min s 2 3 m m k If during turning on no probes are connected the message PROB ERR is displaye...

Page 8: ...ction cannot be transferred to the PC HOLD key It increases the current parameter when used in the menu when used in measurement mode it freezes the measurement in progress and upon application of pre...

Page 9: ...previously been set for the calculation by pressing the FUNC ENTER and UNIT Q T keys When used in the menu it clears or cancels the active function REL key In measurement mode it displays the differe...

Page 10: ...e moment of sampling and will turn off immediately afterwards thus increasing the battery life Using the arrows select YES and confirm using ENTER in order to enable the automatic turning off select N...

Page 11: ...d subsequent stop represent a block When the logging function is on the LOG indication is on the display the battery symbol blinks and a beep is issued each time a logging occurs when using an externa...

Page 12: ...ing on the display upon pressing the SERIAL key is printed the variable associated with the probe connected to the instrument with its relevant unit of measurement or the integral Q t Before starting...

Page 13: ...nu When exceeded these thresholds cause the instrument to stop the integral calculation The probe is detected during turn on and this cannot be performed when the instrument is already on therefore if...

Page 14: ...ressing Start Stop Q t in this status to resume the integration press the key again If one or more limits have been enabled in the menu on reaching of the first the calculation stops Pressing FUNC ENT...

Page 15: ...following table lists all error indications and information displayed by the instrument and supplied to the user in different operating situations Display indications Explanation ERR This appears if t...

Page 16: ...LIMT umolm2 integration limit in mol m2 LOG In PROG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO S...

Page 17: ...the batteries in order to turn the instrument back on In order to avoid data loss the logging session is ended if the HD2102 2 is logging and battery voltage falls below the minimum operating level T...

Page 18: ...Always use good quality leakproof alkaline batteries Sometimes on the market it is possible to find new batteries with an insufficient charge capacity INSTRUMENT STORAGE Instrument storage conditions...

Page 19: ...USB 2 0 connection does not require the setting of parameters The instruments are provided with a complete set of commands and data queries to be sent via the PC All the commands transferred to the i...

Page 20: ...5 10 3600 seconds WC0 Setting SELF off WC1 Setting SELF on Command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combi...

Page 21: ...mand press the MENU key until the _LOG_DUMP_or_ERAS item is displayed and then press on LOG key During data transfer the display shows the message DUMP to stop the data transfer press ESC on the instr...

Page 22: ...g SERIAL again NOTES The print out is formatted across 24 columns Some keys are disabled during serial transmission The following keys work ON OFF HOLD FUNC Max Min Avg and LOG Pressing the HOLD REL a...

Page 23: ...The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and press CONNECT Wait for the connection to occur and follow the indications on the screen F...

Page 24: ...mplete the device will be detected on each new connection automatically In order to check if the entire operation was successful in CONTROL PANEL double click on SYSTEM Select Device Manager and conne...

Page 25: ...ed with instrument off 20 A Mains Output mains adapter 9Vdc 250mA Measuring units lux fcd lux s fcd s W m2 W cm2 J m2 J cm2 mol m2 s mol m2 cd m2 Security of memorized data Unlimited independently of...

Page 26: ...onnector Mains adapter 2 pole connector positive at centre EMC standard regulations Security EN61000 4 2 EN61010 1 level 3 Electrostatic discharge EN61000 4 2 level 3 Electric fast transients EN61000...

Page 27: ...f6 T Temp coefficient 0 05 K Drift after 1 year 1 Working temperature 0 50 C Reference standard CIE no 69 LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equipped with the in...

Page 28: ...9 99 200 0 1999 9 2000 10000 Resolution mol m2 s 0 01 0 1 1 Spectral range 400nm 700nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1...

Page 29: ...9 Resolution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 400nm 1050nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Page 30: ...solution W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 315nm 400nm Peak 360nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fa...

Page 31: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 280nm 315nm Peak 305nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 2 f4 instrument reading error 1digit f5 fatigue...

Page 32: ...on W m2 0 1 10 3 0 001 0 01 0 1 Spectral range 220nm 280nm Peak 260nm Calibration uncertainty 5 f 1 response according to the cosine law 6 f3 linearity 1 f4 instrument reading error 1digit f5 fatigue...

Page 33: ...otted Line Probe Spectral sensitivity LP471 ERY The probe LP 9021 ERY measures the effective total irradiance W m2 eff according to the UV action curve CEI EN 60335 2 27 A particular type of photodiod...

Page 34: ...ngle 2 Measurement range 0 1 cd m2 2000 103 cd m2 LP 471 PAR Quantum radiometric probe for the measurement of the photon flow across the chlorophyll range PAR Photosynthetically Active Radiation 400nm...

Page 35: ...SERIAL INTERFACE AND USB 19 STORING AND TRANSFERRING DATA TO A PERSONAL COMPUTER 21 THE LOGGING FUNCTION ONLY FOR HD2102 2 21 CLEARING THE MEMORY ONLY FOR HD2102 2 21 THE PRINT FUNCTION 22 CONNECTION...

Page 36: ...m d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzentrum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins Einzelne...

Reviews: