![Delta Elektronika SM15K Series Product Manual Download Page 19](http://html1.mh-extra.com/html/delta-elektronika/sm15k-series/sm15k-series_product-manual_2478135019.webp)
ISOLATED ANALOG
SM15K INTERFACES
19 / 19
DELTA ELEKTRONIKA B.V.
rev. Nov. 2019
The Limit Status or LIM-status is "1" in case the output
voltage or current reaches the limit setting. Which limit circuit
is active can be seen on the front panel indicators.
The Over Temperature Status or OT status is "1" in case of
an over temperature, the display will show the OT symbol and
the output shuts down.
The Current Control Status or CC-status output is "1" when
the unit is in CC-mode.
The Power Sink OverLoad Status or PSOL-status is not used
on the SM15K - it is only for SM3300.
The AC-Fail Status or ACF-status output is "1" in case the
input voltage is too low or too high.
The DC-Fail Status or DCF-status output is "1" when the
output voltage is either 5% below or above the set point.
The table in the interface datasheet shows the impedance
levels of the status outputs.
See figure 7 - 7 for a connection diagram.
7.9
REMOTE SHUTDOWN
A voltage of +4 V...+12 V on the Remote ShutDown input on
the programming connector will switch off the output of the
unit.
If the input is high, the RSD icon is shown in the display, see
fig 7 - 8.
It is also possible to use a relay contact or a switch to shut the
unit down (see fig. 7 - 9).
In standby mode the power supply consumes very little
power.
Note: also via Eth or other interfaces RSD can be activated.
7.10
REFERENCE VOLTAGE
A 5.114V reference voltage is available on the connector.
This reference can be used to control the unit with external
potentiometers.
7.11
AUX12V
The 12V on the programming connector can be used to
supply external circuits. The output current is fuse protected
at 0.2A and should not be overloaded.
Output impedance is 5 Ohms.
7.12
TROUBLE SHOOTING
7.12.1
PROGRAMMING DOES NOT WORK OK
Check the programming source is the correct slot.
Check that there is only 1 analog interface installed.
Check the output is switched ON, the Interlock is connected
and Remote ShutDown is low.
7.12.2
STATUS OUTPUT FAIL
Check the wiring of the analog programming connector for
short circuit or loose contacts.
If the front display shows a status icon, but the logic output on
the interface remains low, send the unit back for repair.
7.13
CALIBRATION
The isolated analog interface is factory calibrated.
If desired, both the 5 V and 10 V range can be custom
calibrated. See the ethernet programming manual for more
details.
fig 7 - 8
Position of the RSD icon in the unit display.
fig 7 - 9
Remote ShutDown using a relay contact.