Phantom Installation, Operation & Service Manual
6-8
X-Ray Head Replacement
6.4 X-Ray Head Replacement
Replace x-ray head as follows:
Tools Required:
• None
1
Unplug power supply cord from power outlet.
Warning
Unplug power supply cord before servicing Phantom. Also, do not
perform any maintenance procedures on the unit for at least five
minutes after the power cord has been unplugged. The unit has
large capacitors that need time to discharge their energy after the
unit is unplugged. Failure to follow this can cause serious injury.
2
If not already done so, engage wheel brakes (1 in Figure 6-9).
Figure 6-9. Wheel Brakes
1
Summary of Contents for DynaRad Phantom-8000
Page 6: ...Phantom Installation Operation and Service Manual 4...
Page 24: ...Phantom Installation Operation Service Manual i 18 Applicable Standards...
Page 26: ...Phantom Installation Operation Service Manual ii 2...
Page 66: ...Phantom Installation Operation Service Manual 3 24 Folding Phantom...
Page 74: ...Phantom Installation Operation Service Manual 4 8 Check tube head rotation friction lock...
Page 106: ...Phantom Installation Operation Service Manual 6 22 Release Cable Replacement...
Page 116: ...Phantom Installation Operation Service Manual 8 2 Electrical Schematics...
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 147: ...Phantom Installation Operation Service Manual 9 20 Control Chassis Assembly 501203...
Page 148: ...Phantom Installation Operation Service Manual n 1 Notes...
Page 149: ...Phantom Installation Operation Service Manual n 2 Notes...
Page 150: ...Phantom Installation Operation Service Manual n 3 Notes...
Page 151: ...Phantom Installation Operation Service Manual n 4 Notes...