![Del Medical DynaRad Phantom-8000 Installation, Operation & Service Manual Download Page 30](http://html1.mh-extra.com/html/del-medical/dynarad-phantom-8000/dynarad-phantom-8000_installation-operation-and-service-manual_2465594030.webp)
Phantom Installation, Operation & Service Manual
1-4 Specifications
1.4 Specifications
Table 1-1: Specifications
Specifications
Voltage
104-126VAC/208 - 253 VAC
Line Voltage Regulation
5% maximum at maximum output
Current
30 Amps Maximum Momentary @ 115VAC
Requires 30 amp or greater dedicated breaker or line fuse.
18 Amps Maximum Momentary @ 230VAC
Requires 18 amp or greater dedicated breaker or line fuse.
Frequency
50/60 Hz Single Phase
Generator Type
High frequency inverter, 1.25 kilowatt output
kVp Range
50-100 kVp continuously adjustable, with 1 kVp resolution
mA Range
Fixed, 12.5 mA, constant independent of kVp or time settings
Exposure Time
0.01 to 4.0 seconds in 96 increments
X-Ray Tube
Stationary Anode 100 kVp
Filtration
3.6 mm Aluminum Equivalent or More @ 100kVp
Target Material
Tungsten
Anode Capacity
25,000 Heat Units
Focal Spot
1.0 mm (NEMA)
Beam Angle
15
°
Collimator
Certified Manual
Collimator Lamp Source
30 second timer
Duty
3.3%
Mode of Operation
Momentary
Relative Humidity Limits
Transit/Storage
10% to 100%
Operating
10%-80% Non-Condensing
Atmospheric Limits
14.5 inHg to 30.74 inHg
500 hPa to 1060 hPA
Weight
95 lbs (42 Kg) Total/ X-Ray Generator 39lbs (17.7 kg)/Control
Console (7 lbs (3.2kg)
Certifications:
Classified To UL 187. Certified To
CAN/CSA C22.2 No.
114-M90
.
Equipment not suitable for use in the presence of flammable anesthetic mixtures with air, oxygen or nitrous
oxide.
No user serviceable parts
Summary of Contents for DynaRad Phantom-8000
Page 6: ...Phantom Installation Operation and Service Manual 4...
Page 24: ...Phantom Installation Operation Service Manual i 18 Applicable Standards...
Page 26: ...Phantom Installation Operation Service Manual ii 2...
Page 66: ...Phantom Installation Operation Service Manual 3 24 Folding Phantom...
Page 74: ...Phantom Installation Operation Service Manual 4 8 Check tube head rotation friction lock...
Page 106: ...Phantom Installation Operation Service Manual 6 22 Release Cable Replacement...
Page 116: ...Phantom Installation Operation Service Manual 8 2 Electrical Schematics...
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 147: ...Phantom Installation Operation Service Manual 9 20 Control Chassis Assembly 501203...
Page 148: ...Phantom Installation Operation Service Manual n 1 Notes...
Page 149: ...Phantom Installation Operation Service Manual n 2 Notes...
Page 150: ...Phantom Installation Operation Service Manual n 3 Notes...
Page 151: ...Phantom Installation Operation Service Manual n 4 Notes...