![Del Medical DynaRad Phantom-8000 Installation, Operation & Service Manual Download Page 101](http://html1.mh-extra.com/html/del-medical/dynarad-phantom-8000/dynarad-phantom-8000_installation-operation-and-service-manual_2465594101.webp)
Phantom Installation, Operation & Service Manual
Release Cable Replacement
6-17
7
Attache a feeder wire to the cable sheath. Cut cable (1 in Figure 6-20)
and pull bracket (2) off of cable sheath (3). Discard bracket.
Figure 6-20. Cable and Bracket
8
Make sure the feeder wire is securely attached to the cable sheath.
9
Unscrew two release lever screws (1 in Figure 6-21) and pull release
lever (2) and cable assembly out of unit.
Figure 6-21. Release Lever Screws
2
1
3
1 (x2)
2
Summary of Contents for DynaRad Phantom-8000
Page 6: ...Phantom Installation Operation and Service Manual 4...
Page 24: ...Phantom Installation Operation Service Manual i 18 Applicable Standards...
Page 26: ...Phantom Installation Operation Service Manual ii 2...
Page 66: ...Phantom Installation Operation Service Manual 3 24 Folding Phantom...
Page 74: ...Phantom Installation Operation Service Manual 4 8 Check tube head rotation friction lock...
Page 106: ...Phantom Installation Operation Service Manual 6 22 Release Cable Replacement...
Page 116: ...Phantom Installation Operation Service Manual 8 2 Electrical Schematics...
Page 117: ......
Page 118: ......
Page 119: ......
Page 120: ......
Page 121: ......
Page 122: ......
Page 123: ......
Page 124: ......
Page 125: ......
Page 126: ......
Page 127: ......
Page 147: ...Phantom Installation Operation Service Manual 9 20 Control Chassis Assembly 501203...
Page 148: ...Phantom Installation Operation Service Manual n 1 Notes...
Page 149: ...Phantom Installation Operation Service Manual n 2 Notes...
Page 150: ...Phantom Installation Operation Service Manual n 3 Notes...
Page 151: ...Phantom Installation Operation Service Manual n 4 Notes...