4.4.2.1.3
Display Mode
Determines how the sensor channels will be shown on the display.
INSTRUMENT
DISPLAY MODE
Available selections are:
•
MANUAL
•
AUTO
MANUAL MODE:
If manual mode is selected, the display will default to
the 4 channel summary mode (refer to section 3.2.2.7). Data for a single
channel may be displayed by selecting the desired channel using the
right and left arrow keys:
Pressing either of these keys enables scrolling through and selecting
the active single-channel displays as well as the 4-channel display.
Selection of a channel will display the selected channel continuously
until a different channel is selected.
AUTO MODE:
Automatically scrolls through the available channels,
advancing to the next channel every 3 seconds. In this mode only, the
single-channel displays are shown, the 4-channel summary display is
skipped.
Any selection will automatically revert to the 4-channel summary display
mode during an alarm condition.
4.4.2.1.4 Set Time
INSTRUMENT
SET TIME
Requires numerical entry of the present time in 24-hour clock mode. For
example 1 PM would be entered as 1300. This information is retained
after a power-loss and will not require re-entry.
ENTER TIME AS HHMM
HHMM
4.4.2.1.5 Set Date
INSTRUMENT
SET DATE
Requires numerical entry of the present date in day/month/year format.
ENTER DATE AS DDMMYY
DDMMYY
This information is retained after a power loss and will not require
re-entry. The unit has been designed for compatibility with the year
2000 and will pose no problems when this changeover occurs.
Summary of Contents for Capital Controls CHLORALERT 17CA3000 Series
Page 14: ...325 6610 17 14 Figure 2 1 Receiver Outline Dimensions ...
Page 15: ... 15 325 6610 17 Figure 2 2 Sensor Transmitter Outline Dimensions ...
Page 16: ...325 6610 17 16 Figure 2 3 Interconnection Diagram ...
Page 18: ...325 6610 17 18 Figure 2 4 Sensor Wiring ...
Page 44: ...325 6610 17 44 ...