- 61 - 325.6610.17
Select
EVEN
if parity generation and checking for
EVEN
parity serial byte
protocol
is
required.
If
EVEN PARITY
is not required, select
NONE
.
4.4.2.4.4
Datalink
Enable
DATALINK
ENABLE
ENABLED
The entry in this menu controls whether the instrument is
capable of datalink communications. Allowable selections
are:
•
ENABLE
•
DISABLE
It may be desirable, in some instances, to limit datalink
access to the instrument by disabling the capability of serial
communications. The ability to do this is provided by this
menu.
DATALINK ENABLE
ENABLED
4.4.2.4.5 Datalink
Protocol
COMM
PORT
DATALINK
PROTOCOL
Determines the type of datalink protocol for correct
communication.
Choices
are:
•
STANDARD
•
MODIFIED
STANDARD
selects De Nora Water Technologies “byte-
stuffing”.
MODIFIED
disables datalink “byte-stuffing”.
DATALINK PROTOCOL
STANDARD
4.4.3 Run Calibration
Pressing
ENTER
at the
SETUP-RUN CALIBRATION
menu parameter enables the operator to begin
the
CALIBRATION
procedure described in section 5.0-Calibration.
RUN CALIBRATION
Pressing
ENTER
returns the display to the run mode and causes the unit to begin testing each
channel that is programmed to run the calibration test based on parameters entered in the
SETUP-
CONFIGURE-CHANNELS-CALIBRATION
menu. Two checks are made on each configured
channel:
•
Is calibration gas concentration < 1 ppm?
•
Is calibration gas concentration > Sensor range?
If the answer to either of these questions is yes, the following display will appear:
CALIBRAT'N SETUP ERR
HIT ENTER TO PROCEED
Upon pressing
ENTER
, the unit will return to the
CONFIGURE
menu where the calibration setup
parameters may be corrected by entering the
CONFIGURE-CHANNELS-CALIBRATION
menu.
Refer to section 5.0 for additional calibration information.
Summary of Contents for Capital Controls CHLORALERT 17CA3000 Series
Page 14: ...325 6610 17 14 Figure 2 1 Receiver Outline Dimensions ...
Page 15: ... 15 325 6610 17 Figure 2 2 Sensor Transmitter Outline Dimensions ...
Page 16: ...325 6610 17 16 Figure 2 3 Interconnection Diagram ...
Page 18: ...325 6610 17 18 Figure 2 4 Sensor Wiring ...
Page 44: ...325 6610 17 44 ...