1.3 Specifications
SYSTEM:
Gases
Chlorine
Sulfur
Dioxide
Channels/Gases
4 max., channels and gas interchangeable
for each receiver
Gas Conc. Ranges
Chlorine: 0-5, 0-10, 0-50 ppm
Sulfur Dioxide: 0-10, 0-20, 0-100 ppm
Ambient Temperature
Receiver: -5 to +55°C (23 to 131°F)
Sensor: -30 to +55°C (-22 to 131°F)
Ambient Pressure:
0.8 to 1.2 atmospheres
AC POWER:
Voltage
Range
93.5 to 276 Vac
Frequency Range
47-63 Hz
Power Consumption
Receiver- 10 watts nominal, 12 watts max.
Electrochemical Sensor/Transmitter- 0.5 watts max.
Voltage Dropout Duration
< 20 ms
Permitted
RECEIVER:
Signal
Digital Communication
RS 232/485 (optional),
4-20 mA each channel (optional)
Relays
3 SPDT Standard, Additional 6 SPST optional
Relay contacts rated at 3.15 Amps/240 Vac
Optional relays are field replaceable
Annunciators
Flashing display indication for concentration alarms and other
significant events. Display backlight turned off during battery
operation or if battery is faulty.
Alarm Levels
Selectable, 3 programmable concentration alarm levels
for
each
channel.
Alarm Modes
Selectable, TWA, Instantaneous, STEL, and Customer Defined
Alarm Operation
Selectable, normal (self-resetting), latched and delayed.
Display
Back-lit LCD, 2 lines x 20 characters
Keyboard
Polyester, 20 tactile key position
Diagnostics
System, power and sensor failure
Security
Password-protected
Summary of Contents for Capital Controls CHLORALERT 17CA3000 Series
Page 14: ...325 6610 17 14 Figure 2 1 Receiver Outline Dimensions ...
Page 15: ... 15 325 6610 17 Figure 2 2 Sensor Transmitter Outline Dimensions ...
Page 16: ...325 6610 17 16 Figure 2 3 Interconnection Diagram ...
Page 18: ...325 6610 17 18 Figure 2 4 Sensor Wiring ...
Page 44: ...325 6610 17 44 ...