7.

 

Configuration 

7.1

 

Transmitter Configuration 

This section mainly introduces how to use the 3 local keys (Enter key←┘), (up key "↑"), (down key 

"↓"), combined with LCD display to configure internal parameters. 

 

7.1.1 LCD display function overview 

See (Figure 7-1) for full screen information. The variables that can be displayed on the LCD screen 

include current (Figure 7-2), percentage (Figure 7-3) and main variable (Figure 7-4). You can set to 

display a single variable, or you can set the percentage and the main variable to cycle display, and 

the effective digits of the decimal part can be set (0, 1, 2, 3, 4 decimal places). The display mode is 

set  by  475  configuration  software  or  button.  If  the  percentage  and  main  variable  are  set  to  be 

displayed cyclically, the LCD will display these two variables alternately with a period of 3 seconds. 

 

Figure 7-1 Full LCD display                                                  Figure 7-2 Current display 

 

Figure 7-3 Percentage display                                      Figure 7-4 Main variable display 

 

7.1.2 LCD Display Symbol Description 

①    If it is in HART communication status, the symbol 

  on the right corner of the LCD 

screen will flash. 

Summary of Contents for UHC Magnetic Level Gauge

Page 1: ...1 17 WWW DDTOP COM UHC Bb Remote Transmission MAGNETIC LEVEL GAUGE Operating Manual UHC Bb DT JS 1027 2018 A...

Page 2: ...and are subject to change without notice Dandong Top Electronics Instrument Group Co Ltd reserves the right to modify the products described in this manual at any time without prior notice Quality As...

Page 3: ...arranty Quality Dandong Top Electronics Instrument Group Co Ltd has passed the ISO9001 quality system certification The whole process of product production is strictly in accordance with the scope of...

Page 4: ...Product Descriptions 5 3 Technical Parameters 6 4 Outline Dimension Diagram 7 5 Unpacking and Inspection 8 6 Installation 8 7 Configuration 10 8 Precautions 15 9 Fault Analysis and Troubleshooting 16...

Page 5: ...onal standards codes and regulations Ensure that the equipment is installed in accordance with intrinsically safe or non flammable site operating procedures 1 2 Process leakage may cause serious injur...

Page 6: ...y or output to the control room through the output terminal to achieve long distance transmission of liquid level 2 3 Packaging Please take the packaging waste to a special recycling facility 2 4 Lift...

Page 7: ...40 80 Working Temperature 40 200 When T 150 the transmitter does not allow heat preservation Accuracy 5mm Power Inlet M20 1 5 female thread or as customer requirement Explosion proof Type Intrinsicall...

Page 8: ...product nameplate is consistent with the packing list information Figure 4 Figure 4 Product Nameplate 5 1 2 Check the quantity and material of each part against the packing list 5 2 Content of Inspect...

Page 9: ...point full scale of the remote transmission device marked at the factory and the O scale full scale of the on site indicator on the measuring tube are one to one correspondence and then fixed with a...

Page 10: ...lay a single variable or you can set the percentage and the main variable to cycle display and the effective digits of the decimal part can be set 0 1 2 3 4 decimal places The display mode is set by 4...

Page 11: ...n the 888 character in the lower left corner shows 1 17 it indicates that the transmitter is in the field configuration mode with menu rotation At this time you can input the initialization code modif...

Page 12: ...the key the up and down arrows on the left will light up indicating that the setting is saved 7 Press again to restart setting data 7 1 5 Key function code quick reference table When using button conf...

Page 13: ...n arrows on the left will light up Unit Setup 7 1 7 Range setting Set the lower limit of the range LRV Long press the key to enter the configuration mode first menu No 1 is displayed the up and down a...

Page 14: ...ation First adjust the PV value to the zero position Move the magnetic float to zero Switch the menu to No 11 and the LCD will display the current calibration position value at the same time and then...

Page 15: ...confirm and start the next digit setting and set other digits and decimal points in sequence When the up and down arrows on the left side light up the menu function setting is complete Compensation v...

Page 16: ...n the container etc Please refer to the instructions in the chapter 6 2 Installation Operation Process and 7 1 Electrical Wiring and remove the components in the reverse order of the operation steps 1...

Page 17: ...17 17 WWW DDTOP COM...