menu item setting is completed when the up and down arrows on the left light up.
Set the upper limit URV
Switch the menu to No. 4, and the LCD displays the current URV value at the same time, and then
repeat the steps ③ and ④ in 7.1.6 to set.
Range Setting
7.1.8 Damping setting
Switch the menu to No. 5, and the LCD will display the current damping value at the same time, and
then repeat the steps ③ and ④ in 7.1.6 to set until the completion.
7.1.9 Two-point calibration
LRV calibration
①
First adjust the PV value to the zero position (Move the magnetic float to zero)
②
Switch the menu to No. 11, and the LCD will display the current calibration position value at the
same time, and then repeat ③④ in 7.1.6 to input the zero point value for zero point position
calibration.
③ Press the "←┘" key until the up and down arrows on the left light up the zero position
calibration is completed
URV calibration
②
First adjust the PV value to the full-scale position (Move the magnetic float to zero)
③
Switch the menu to No. 12, and the LCD will display the current calibration position value at
the same time, and then repeat ③④ in 7.1.6 to input the full scale value for full scale position
calibration