The Model IL70A is a single-channel instrument which is easily configured, rugged, analog based 
signal conditioner made to operate with traditional Wheatstone mv/V stain gage bridge sensors 
used to measure force, pressure, torque and other DC strain gage related measurements. The Model 
IL70A has an extremely wide DC power input range with user selectable gain, zero and anaolg output 
making it adaptable to any users sensor application that requires stable, repeatable and noise free 
analog signals for control or data acquisition.  

1

G

ENERAL

D

ESCRIPTION AND

S

PECIFICATIONS

IN-LINE DC Strain Gage Conditioner

M

ODEL 

IL70A

DC Strain Gage

C

ONDITIONER 

Module

IL70A

GENERAL DESCRIPTION AND SPECIFICATIONS  IL70A.1

Top View of the Model IL70A DC Strain Gage Module (cover removed)

DC Strain Gage Symmetry Operation:

 

For applications requiring precision in both tension and compression (CW & CCW for 

torque cells), the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer. 

The Model IL70A is best calibrated by means of a “two-point (dead-weight)” or shunt-calibration 
technique, which is outlined in section three.  The supplied calibration resistor is 59 kilohms, 1% 
which is located internally to the IL70A 

pcb circuit.

 

Shunt

  Operation:

 

Located  between  the  two  screw  terminal  connectors,  the  IL70A  contains  a  Shunt  resistor  jumper 

selection.  "EXT" is for an external - user installed shunt or "INT" for the internal - built in - 59K Ohm shunt. Installing the jumper 

enables the selected feature. For an external shunt resistor, install the leaded resistor between the "Cal Sense" (pin 7) terminal 

and either the + Signal (pin 5) or the - Signal (pin 6) terminals. Use the SW4 "Shunt" button to engage the shunt resistor across 

the selected side of the bridge or "SHP or "SHN" logic input lines. For transducers with an installed shunt resistor, place the 

jumper on the "EXT" teminals and wire the "Cal Sense" line to the transducer's shunt connector pin.

9237

5

.00 rev. 

B

Summary of Contents for IL70A

Page 1: ...CCW for torque cells the IL70A contains a symmetry adjustment which corrects for negative span or slope of the transducer The Model IL70A is best calibrated by means of a two point dead weight or shun...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...henden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse a...

Page 6: ...ble up to 70mA Power Supply 11 28 Vdc regulated 2 watts max Analog Ouput selectable 0 to 5 0 to 10 Vdc 4 12 20 or 4 20mA 20 over range voltage only Operating Temperature 10 to 70 Degrees C 5 to 95 rel...

Page 7: ...Label Function 1 2 3 4 5 6 7 8 Analog Output SHP SHN NC Analog Common Power Common 24 Vdc Shield Signal Output Shunt Positive switch input Shunt Negative swith input No connection Signal Output Power...

Page 8: ...located between the two wiring connectors Below are the appropriate wiring and jumper selections depending on the application using the recommended 7 wire configuration TRANSDUCER CONNECTIONS IL70A 2...

Page 9: ...st the instrument sensitivity accordingly Place the jumper on the IL70A to INT to enable the 59K Ohm resistor A one percent 59 kilohm calibration resistor is installed in the IL70A at the factory The...

Page 10: ...o obtain your nominal full scale analog output Use Fine Span control to adjust the signal to the precise value required or for an analog value that correspondes with the appropriate Shunt Calibration...

Page 11: ...l The Remote Cal function provides a convenient method of periodically monitoring calibration of the instrument in the positive or negative realm Switch 1 Mode current I or voltage analog output Switc...

Page 12: ...by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output readstotally off scale and the BALANCE controls have no authority can very likely...

Page 13: ...Fig 4 Star Bridge DC Strain Gage Conditoner Module IL70A...

Page 14: ...oduct or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPAIRED OR REPLACED UNDER WARRANTY MUST BE RETURNED TO THE FACTORY Once the customer...

Reviews: