4. VERIFICATION OF NORMAL OPERATION

It is the purpose of this section to aid the user in determining, in the event of a

malfunction to which the Model 5M78 is suspected of contributing, whether the

instrument is functioning normally or whether it is the source of the observed

trouble. In the event the unit requires repair, the user may also contact the 

factory Service Department or the local Daytronic Representative for assistance.

Daytronic service information is located on the last page of this document.

If the instrument is suspected of faulty operation, follow the steps below.

(a)

   

 

  

If the unit is totally inoperational (front-panel power indicator does not

light), check the primary power input terminals for proper connection.

Input power can be from 11 to 28 Vdc and will draw less than 2 watts.

If properly connected, the front panel Green LED will be illuminated. Before

reapplying power, visually inspect the power supply and the input power

connections for any discrepancy which could have caused the overload.

(b) 

   

  

 

If the transducer has some preloading, the BALANCE controls may not

allow successful zeroing of the instrument output. This condition can be

remedied by connecting a resistor (50 k- 200 k range, metal-film type)
from the +Signal terminal of the transducer to the + or –

Excitation Sense

terminals. The 

Excitation 

terminal to which the connection is made is

determined by the direction of the loading or off-zero reading.

(c) 

   

The inability to balance correctly, where the instrument output reads

totally off scale and the BALANCE controls have no authority, can very

likely be the result of a damaged or defective transducer or cable. This

possibility can be confirmed (or eliminated) by substituting a transducer

and cable known to be in good condition or by simulating a balanced

transducer, using either a commercially available transducer simulator

or the simple 

star bridge

 arrangement shown in Figure 7. The 

star bridge

simulates a conventional four-arm bridge in an exact condition of bal-

ance. To construct a 

star bridge, 

connect four 10% carbon resistors as

shown in Figure 5. Use 180-ohm resistors to simulate a 350-ohm bridge

and use 56-ohm resistors to simulate a 120-ohm bridge. Neither the

resistor values nor temperature characteristics are critical since the

balance condition of a star bridge is not determined by the resistance 

Figure 7. Star Bridge Construction

20

values. Solder two resistors together, then solder the remaining two

resistors together. Next, connect the two junctions together using a

separate wire as shown. There is a good reason for this method of

construction, and it should be followed. Connect the substitute or simu-

lated transducer to the instrument I/O connector using a short 4-wire

cable configuration as shown in Figure 4. Attempt to balance the substitute

simulated transducer. If conditions now appear to be normal, the transducer

or cable is at fault. If the previous difficulties persist, the 5M78 is faulty.

Summary of Contents for 5M78

Page 1: ...the 5M78 by gently pulling the clear plastic cover from the bottom side so the cover rotates open from DIN AC Strain GageConditioner MODEL 5M78 AC Strain Gage CONDITIONER Module 5M78 1 A 1 5M78 GENERA...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...achenden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse...

Page 6: ...al value 0 08 sec 0 008 sec 100 Hz 0 0011 sec 650 Hz To 0 1 of final value 0 095 sec 0 0095 sec 100 Hz 0 0016 sec 650 Hz To 0 02 of final value 0 10 sec 0 01 sec 100 Hz 0 0014 sec 650 Hz 5M78 1 GENERA...

Page 7: ...SENSE Top Front 4 4 SHIELD Bottom Front 1 1 Ano Out ANALOG Output Bottom Front 2 2 SHP Shunt Positive Bottom Front 3 3 SHN Shunt Negative n Bottom Front 4 4 SYC SYNC MASTER SLAVE TRANSDUCER CONNECTION...

Page 8: ...ration resistor for convenience in subsequent checking Replace the installed calibration resistor with one or an equivalent resistance value supplied by the transducer manufacturer to achieve a precis...

Page 9: ...ent dead weight value greater than one half of full scale Adjust Coarse Span for a nominal corresponding output Adjust the PHASE control until a maximum output value is obtained Once set for your tran...

Page 10: ...ted on the calibration document The SYMMETRY control has approximately 7 authority full scale If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent...

Page 11: ...Switch 2 Volts selects 5 or 10 Vdc when mode is voltage Switch 3 Current selects 4 12 20 or 4 20mA when mode is current Switch 4 Filter selects 650 Hz or 100 Hz at 3 db for 100 Hz switch 5 must be se...

Page 12: ...made is determined by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authori...

Page 13: ...Contact Switch NO NOT PUSHED Logic 1 Shunt not enabled PUSHED Logic 0 Shunt is enabled Push Button NO 11 28 VDC POWER SUPPLY SHIELD Fig 3 Shunt Connections Fig 4 Star Bridge...

Page 14: ...urment setup beat frequencies can be produced by the excitation clock oscillator To prevent this the user should declare one of the 5M78 units as a Master via switch 6 on the front panel The other uni...

Page 15: ...orkmanship or the refund of the purchase price to the Buyer Before returning a product or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPA...

Page 16: ...Daytronic Corporation Dayton OH USA w ww ww w d da ay yt tr ro on ni ic c c co om m...

Reviews: