3 GENERAL DESCRIPTION AND CALIBRATION

      The Model 5M78 is useful in applications involving tranformer coupling to the strain gage

bridge (for example, rotary transformer torque sensors) and in certain appications that require high  
sensitivity with optimum signal-to-noise characteristics. Carrier amplifiers offer higher sensitivity
than dc amplifiers and, since they respond only to the modulated carrier frequency, they reject certain 
extraneous voltages that can cause error in dc systems. These error sources include thermocouple 

or galvanic voltages in the cable-connector system, homopolar voltages from rotating machinery,
low-frequency pickup, and 1/f noises of various origin. 

       

Calibration of the 5M78

     is accomplished by the conventional shunt technique, using an internally 

(or externally) installed calibration resistor, or via the "dead-weight" method. 

Calibration Resistor.   

If a fixed resistor is shunted across one arm of a strain

gage bridge, it produces an unbalance equivalent to that of a particular value of

mechanical input. If this Equivalent Input value is accurately known, it can be used

as a reference point for shunt calibration of the system. Upon completion of

installation of the transducer and its associated cabling, the user can:

Perform an overall dead

 

weight calibration using a precisely known

value of mechanical input. The calibration can then be transferred to the

installed calibration resistor for convenience in subsequent checking.

Replace the installed calibration resistor with one (or an equivalent

resistance value) supplied by the transducer manufacturer to achieve a

precisely known Equivalent Input allowing the instrument sensitivity to

be adjusted correctly.

 

Determine the Equivalent Input value for the installed calibration resis-

tor, knowing the transducer sensitivity, and adjust the instrument sen-

sitivity accordingly.

A one percent 59-kilohm calibration resistor is installed in the 5M78 at the factory.

The installed resistor can usually be used even though the transducer calibration

data mentions some other resistance value. In Section 4 of this manual, the tech-

niques described above are demonstrated. If, however, the installed value of the

calibration resistor is not appropriate for the transducer and measurement range to

be used, the 59-kilohm resistor should be replaced at this time. The calibration

resistor is mounted on terminals located internally to the 5M78 conditioner's

printed-circuit board. It can be accessed by removing the instrument case (Fig. 5).

Note:

 

A variety of Rotary Transformer Torque Transducers are supplied

with the appropriate calibration resistor integral to the transducer. When this type

of transducer is used it will be necessary to place a short across the 59-kilohm

resistor internal to the instrument. The transducer calibration resistor can be appropri-

ately connected to the 5M78 calibration circuit via the transducer cabling. Refer to

cabling diragam section of this manual in the area of CAL SENSE.

7

(1)

(2)

(3)

Calibration  5M78.3

Summary of Contents for 5M78

Page 1: ...the 5M78 by gently pulling the clear plastic cover from the bottom side so the cover rotates open from DIN AC Strain GageConditioner MODEL 5M78 AC Strain Gage CONDITIONER Module 5M78 1 A 1 5M78 GENERA...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...achenden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse...

Page 6: ...al value 0 08 sec 0 008 sec 100 Hz 0 0011 sec 650 Hz To 0 1 of final value 0 095 sec 0 0095 sec 100 Hz 0 0016 sec 650 Hz To 0 02 of final value 0 10 sec 0 01 sec 100 Hz 0 0014 sec 650 Hz 5M78 1 GENERA...

Page 7: ...SENSE Top Front 4 4 SHIELD Bottom Front 1 1 Ano Out ANALOG Output Bottom Front 2 2 SHP Shunt Positive Bottom Front 3 3 SHN Shunt Negative n Bottom Front 4 4 SYC SYNC MASTER SLAVE TRANSDUCER CONNECTION...

Page 8: ...ration resistor for convenience in subsequent checking Replace the installed calibration resistor with one or an equivalent resistance value supplied by the transducer manufacturer to achieve a precis...

Page 9: ...ent dead weight value greater than one half of full scale Adjust Coarse Span for a nominal corresponding output Adjust the PHASE control until a maximum output value is obtained Once set for your tran...

Page 10: ...ted on the calibration document The SYMMETRY control has approximately 7 authority full scale If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent...

Page 11: ...Switch 2 Volts selects 5 or 10 Vdc when mode is voltage Switch 3 Current selects 4 12 20 or 4 20mA when mode is current Switch 4 Filter selects 650 Hz or 100 Hz at 3 db for 100 Hz switch 5 must be se...

Page 12: ...made is determined by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authori...

Page 13: ...Contact Switch NO NOT PUSHED Logic 1 Shunt not enabled PUSHED Logic 0 Shunt is enabled Push Button NO 11 28 VDC POWER SUPPLY SHIELD Fig 3 Shunt Connections Fig 4 Star Bridge...

Page 14: ...urment setup beat frequencies can be produced by the excitation clock oscillator To prevent this the user should declare one of the 5M78 units as a Master via switch 6 on the front panel The other uni...

Page 15: ...orkmanship or the refund of the purchase price to the Buyer Before returning a product or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPA...

Page 16: ...Daytronic Corporation Dayton OH USA w ww ww w d da ay yt tr ro on ni ic c c co om m...

Reviews: