Remote Calibration Check

   The instrument can be placed in the calibration mode

(positive or negative) by shorting terminals Power Common and SHP for positive or SHN for

negative on the lower I/O connector. Figure 4 indicates two methods of remotely

entering the calibration mode (external switch, or push button control). The

Remote Cal function provides a convenient method of periodically monitoring

calibration of the instrument in the positive or negative realm.

Master/Slave Connections.

 

    When more than one 5M78 is being used in a measurement

setup (instruments are closely mounted or the transducer cabling is in a common conduit

or raceway), beat frequencies may be produced from the 3-kHz oscillators used in

the instruments to develop the excitation. To prevent beat frequencies from occur-

ring, one unit can be designated the master, and the remaining units can be driven

from the oscillator contained in the master unit. The remaining units are designated

as slave instruments. To perform master/slave wiring, refer to Figure 7.

3.

CALIBRATION 

(cont.)

Switch 1 - Mode

 - selects current (I) or voltage analog output

Switch 2 - Volts

  - s/- 5 or +/- 10 Vdc when mode is voltage

Switch 3 - Current

 - selects 4-12-20 or 4-20mA when mode is current

Switch 4 - Filter

 - selects 650 Hz or 100 Hz at 3 db, for 100 Hz switch 5 must be set to the right 

Switch 5 - Filter

 - selects 10 Hz or 100 Hz at 3 db, for 100 Hz switch 4 must be set to the right 

Switch 6 - Sync

 - selects excitaton clock to slave from a master module or to be master 

Switch 7 - Zero Adj.

 - selects extended (100%) or normal (25%) zero authority 

Switch 8 - Aux Output

 - selects SYNC terminal output for use by other linked modules

Coarse Zero

 - 16 position switch adjustment for stepped zero balance control

marked for calibration default position

 

Fine Zero

 - 18 turn potentiometer for fine zero balance control

Coarse Span

 - 16 position switch adjustment for stepped gain contr ol

marked for calibration default position

 

Fine Span

 - 18 turn potentiometer for fine gain - span control

Symmetry 

-  adjust the negative output span to be equal to the positive output 

Phase 

-  adjustment for phasing the AC modulated signal to its highest level at load 

Over Range 

- indicates when the analog output is 2% greater than mode selected 

Power

 - indicates the power input voltage is ON

Fig. 2 Front Panel Settings and Indicators

Calibraton 5M78.3 

or in the AUX position 

for raw - uncalibrated

 analog

 gained

 voltage output

Summary of Contents for 5M78

Page 1: ...the 5M78 by gently pulling the clear plastic cover from the bottom side so the cover rotates open from DIN AC Strain GageConditioner MODEL 5M78 AC Strain Gage CONDITIONER Module 5M78 1 A 1 5M78 GENERA...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...achenden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse...

Page 6: ...al value 0 08 sec 0 008 sec 100 Hz 0 0011 sec 650 Hz To 0 1 of final value 0 095 sec 0 0095 sec 100 Hz 0 0016 sec 650 Hz To 0 02 of final value 0 10 sec 0 01 sec 100 Hz 0 0014 sec 650 Hz 5M78 1 GENERA...

Page 7: ...SENSE Top Front 4 4 SHIELD Bottom Front 1 1 Ano Out ANALOG Output Bottom Front 2 2 SHP Shunt Positive Bottom Front 3 3 SHN Shunt Negative n Bottom Front 4 4 SYC SYNC MASTER SLAVE TRANSDUCER CONNECTION...

Page 8: ...ration resistor for convenience in subsequent checking Replace the installed calibration resistor with one or an equivalent resistance value supplied by the transducer manufacturer to achieve a precis...

Page 9: ...ent dead weight value greater than one half of full scale Adjust Coarse Span for a nominal corresponding output Adjust the PHASE control until a maximum output value is obtained Once set for your tran...

Page 10: ...ted on the calibration document The SYMMETRY control has approximately 7 authority full scale If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent...

Page 11: ...Switch 2 Volts selects 5 or 10 Vdc when mode is voltage Switch 3 Current selects 4 12 20 or 4 20mA when mode is current Switch 4 Filter selects 650 Hz or 100 Hz at 3 db for 100 Hz switch 5 must be se...

Page 12: ...made is determined by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authori...

Page 13: ...Contact Switch NO NOT PUSHED Logic 1 Shunt not enabled PUSHED Logic 0 Shunt is enabled Push Button NO 11 28 VDC POWER SUPPLY SHIELD Fig 3 Shunt Connections Fig 4 Star Bridge...

Page 14: ...urment setup beat frequencies can be produced by the excitation clock oscillator To prevent this the user should declare one of the 5M78 units as a Master via switch 6 on the front panel The other uni...

Page 15: ...orkmanship or the refund of the purchase price to the Buyer Before returning a product or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPA...

Page 16: ...Daytronic Corporation Dayton OH USA w ww ww w d da ay yt tr ro on ni ic c c co om m...

Reviews: