3. CALIBRATION 

(cont.)

This section contains the instructions for calibrating the 5M78. Included is a

functional description of the instrument front-panel (see Figure 2). To perform

calibration, proceed as follows.

Turn power ON to the 5M78 DC Power input terminal (11 to 28 Vdc input)

The front-panel indicator should light green to indicate the application of

dc power. Allow 10 minutes of warmup for stabilization of transducer

characteristics. Open the protective, clear plastic, front cover of the unit.

Set the Coarse Zero and Span controls to the marked label position (

<

, Min

>

).

Position the front panel switches to the desired settings for the application.

Refer to Figure 2 for details. If using Shunt method, refer to step (k).

 

(b)

(a)

(c)

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

With the transducer unloaded, adjust the Coarse Zero control and the Fine Zero

control until the desired analog output is achieved. If greater zero authority is

 desired, place the front panel Switch 7 -  "Zero Adj"  to the extended range. This will allow

the user to obtain 100% zero offset control. The Normal postion of switch 7 allows 

for approximately 25% Zero authority. 

(d)

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

Load the transducer in the positive direction with a convenient dead weight value 

(greater than one half of full scale). Adjust Coarse Span for a nominal corresponding output.

 Adjust the PHASE control, until a maximum output value is obtained. (Once set for your 

transducer, this PHASE ADJUSTMENT step need not be repeated unless a change in

cable length – or transducer – is required). Unload the sensor and re-Zero as needed. 

(e)

(f)                  If the calibration will utilized a known "dead-weight", re-load the sensor and adjust the

Coarse and Fine Span controls for the expected analog output value. Unload - Zero and

Load - Span operations may need to be repeated for analog output precision.

Note: With sensor unloaded, activate & record "SHP"  value for furture calibration verification.

If the calibration will utilize the "Shunt" calibration method, refer to step (k).

recommended calibration resistor.

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(g)

(m)

Calibration 5M78.3

 

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(m)

If the transducer is to be also used in the negative realm, load the

transducer in the negative direction with the same dead weight value as

used in step (f) and confirm that the dead weight reading obtained is the

same, but negative polarity, as that of step (f). If not, see steps (i) thru (j).

A Symmetry adjustment is provided to compensate for transducers

that do not have symmetrical sensitivity characteristics. This

adjustment is factory set assuming symmetrical characteristics. If step

(h) indicates that a field adjustment is necessary, proceed as follows.

recommended calibration resistor.

the Coarse and Fine SPAN controls. If not, adjust the Symmetry control

to obtain the negative Equivalent Input value.

If dead weight calibration is not practical and the transducer calibration

data is unknown, the Equivalent Input value for the factory-installed

calibration resistor can be approximated

mv/v sensitivity rating of the transducer

known.

as follows, assuming that

and the bridge resistance

the

are

13

(h)

(i)

(m)

Summary of Contents for 5M78

Page 1: ...the 5M78 by gently pulling the clear plastic cover from the bottom side so the cover rotates open from DIN AC Strain GageConditioner MODEL 5M78 AC Strain Gage CONDITIONER Module 5M78 1 A 1 5M78 GENERA...

Page 2: ...miento de este instrumento debe ser realizada por personal calificado solamente El C digo El ctrico Nacional define a una persona calificada como una que est familiarizada con la construcci n y operac...

Page 3: ...r elektrische Sicherheitsauflagen f r Angestellt Arbeitspl tze ANSI NFPA 70E 2012 der Vereinigten Staaten und alle zus tzlichen Arbeitsplatzsicherheitsauflagen folgen die auf Ihre Installation anwend...

Page 4: ...y que el material aislante no tenga rajaduras Reemplace de inmediato cualquier parte defectuosa Mesures de S curit Les mesures de s curit suivantes doivent tre prises chaque fois qu un type de connex...

Page 5: ...achenden Schaltung hergestellt werden Tragen Sie Schutzbrillen und Isolierhandschuhe wenn Sie Anschl sse an den Stromkreisen vornehmen H nde Schuhe und Fu boden m ssen trocken sein wenn Sie Anschl sse...

Page 6: ...al value 0 08 sec 0 008 sec 100 Hz 0 0011 sec 650 Hz To 0 1 of final value 0 095 sec 0 0095 sec 100 Hz 0 0016 sec 650 Hz To 0 02 of final value 0 10 sec 0 01 sec 100 Hz 0 0014 sec 650 Hz 5M78 1 GENERA...

Page 7: ...SENSE Top Front 4 4 SHIELD Bottom Front 1 1 Ano Out ANALOG Output Bottom Front 2 2 SHP Shunt Positive Bottom Front 3 3 SHN Shunt Negative n Bottom Front 4 4 SYC SYNC MASTER SLAVE TRANSDUCER CONNECTION...

Page 8: ...ration resistor for convenience in subsequent checking Replace the installed calibration resistor with one or an equivalent resistance value supplied by the transducer manufacturer to achieve a precis...

Page 9: ...ent dead weight value greater than one half of full scale Adjust Coarse Span for a nominal corresponding output Adjust the PHASE control until a maximum output value is obtained Once set for your tran...

Page 10: ...ted on the calibration document The SYMMETRY control has approximately 7 authority full scale If dead weight calibration is not practical and the transducer calibration data is unknown the Equivalent...

Page 11: ...Switch 2 Volts selects 5 or 10 Vdc when mode is voltage Switch 3 Current selects 4 12 20 or 4 20mA when mode is current Switch 4 Filter selects 650 Hz or 100 Hz at 3 db for 100 Hz switch 5 must be se...

Page 12: ...made is determined by the direction of the loading or off zero reading c The inability to balance correctly where the instrument output reads totally off scale and the BALANCE controls have no authori...

Page 13: ...Contact Switch NO NOT PUSHED Logic 1 Shunt not enabled PUSHED Logic 0 Shunt is enabled Push Button NO 11 28 VDC POWER SUPPLY SHIELD Fig 3 Shunt Connections Fig 4 Star Bridge...

Page 14: ...urment setup beat frequencies can be produced by the excitation clock oscillator To prevent this the user should declare one of the 5M78 units as a Master via switch 6 on the front panel The other uni...

Page 15: ...orkmanship or the refund of the purchase price to the Buyer Before returning a product or products for any reason the customer must call Daytronic Customer Support Services at ALL EQUIPMENT TO BE REPA...

Page 16: ...Daytronic Corporation Dayton OH USA w ww ww w d da ay yt tr ro on ni ic c c co om m...

Reviews: