Setup Menu
26
1. Choose Set Temp Cal from the Setup menu.
The Set Temp Cal dialog box appears. The raw reading shown in this dialog
box indicates the unadjusted reading coming from the sensor. The adjusted
reading indicates what the software displays, adjusted for any previously
entered calibration number.
S
ET
T
EMP
. C
AL
2. Enter the desired adjusted reading and choose OK.
The software automatically calculates the necessary calibration number to
convert the raw reading into the adjusted reading and saves that calibration
number to the console and the station’s configuration file.
Set Hum Cal
If you need to calibrate your outside humidity reading, you must set the calibra-
tion number on your station console using the software. For performance reasons,
the software reads the calibration number from the station’s configuration file,
rather than reading it from the console itself (see “Calibration Numbers” on
page 93). Therefore, in order for the software to display the correct outside
humidity, you must set the outside humidity calibration number from the software.
Note:
Consult your station manual for a more detailed description of the outside humidity calibra-
tion number.