SPECIFICATIONS
1.1
General specifications
Module size, W
×
D
×
H
75
×
180
×
112 mm/3.0
×
7.1
×
4.4 in
Module weight
0.7 kg/1.5 lbs
Power consumption
about 4 W
1.2
Typical performance
Oscillometric measurement principle.
Measurement range
adult
25...260 mmHg
child
25...195 mmHg
infant
15...145 mmHg
Accepted HR
30...250 bpm
Measurement interval
from continuous to 1h, 2h, 4h
Measurement time, typical
adult
23 s
infant
20 s
Initial inflation pressure
adult
185
±
10 mmHg
child
150
±
10 mmHg
infant
120
±
10 mmHg
Venous stasis
adult
80
±
10 mmHg / 2 min
child
60
±
10 mmHg / 2 min
infant
40
±
10 mmHg / 1 min
Cuff widths
Please see
User’s Guide
1.3
Technical specifications
Deflation rate, HR dep.
5...13 mmHg/sec
Inflation rate, typical
20...185 mmHg, 1...5 s
Automatic software control, max. inflation pressure
adult
280
±
10 mmHg
child
200
±
10 mmHg
infant
150
±
10 mmHg
Over pressure limit, stops measurement after 2 seconds
adult
320 mmHg
child
220 mmHg
infant
165 mmHg
Safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg in
infant mode. Independent timing circuit limits pressurizing (>5 mmHg) time to 2 minutes 10
seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
Zeroing to ambient pressure is done automatically.
Summary of Contents for S/5 M-NIBP
Page 2: ......
Page 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Page 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Page 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Page 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...