NIBP Module, M-NIBP
Document No. 800 1015-1
7
2.3
Connectors and signals
2.3.1
Module bus connector
13
1
14
25
Module Bus connector (X1)
Pin No
I/O
Signal
Not used
1
I
RESET_RS485
2
I
-15 VDC
3
I
+15 VDIRTY
4
I
+15 VDC
5
I/O
-DATA_RS485
6
I/O
DATA_RS485
7
-
Ground & Shield
8
I
-RESET_RS485
9
I
CTSB
*
10
O
RTSB
*
11
I
RXDB
*
12
O
TXDB
*
13
-
Ground & Shield
14
I
+32 VDIRTY
*
15
I
GroundDIRTY
16
I
CTSC
*
17
O
RTSC
*
18
I
RXDC
*
19
O
TXDC
*
20
-
ON/STANDBY
*
21
-
PWM_ECG
*
22
-
RXDD_RS232
*
23
-
TXDD_RS232
*
24
I
+5 VDC
25
I
+5 VDC
*Not used in the M-NIBP module