![Datex-Ohmeda S/5 M-NIBP Technical Reference Manual Slot Download Page 25](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-nibp/s-5-m-nibp_technical-reference-manual-slot_513844025.webp)
NIBP Module, M-NIBP
Document No. 800 1015-1
21
4.2
NIBP error code explanation
Code
Explanation
0
RAM failure
Memory failure. Change NIBP board.
1
ROM checksum error
Memory failure. Change NIBP board.
2
+15V failure
Check short circuits. Change NIBP board.
3
-15V failure
Check short circuits. Change NIBP board.
4
EEPROM protection switch error. (only with S-STD93)
Turn the toggle switch to the left at the bottom of the module.
5
Calibration not protected. (only with S-STD93)
Protect calibration by selecting Protection ON in the NIBP calibration menu
6
ADC error
ADC circuit failure. Change NIBP board.
7
Watchdog time too short
Change NIBP board.
8
Watchdog time too long
Change NIBP board.
9
Watchdog activated
Change NIBP board.
10
EEPROM checksum error
Memory failure. Change NIBP board.
11
Auto zero range exceeded
Calibrate NIBP.
12
Communication break
Temporal break down of communication from monitor detected. Automatic
recovery.
14
Too early Auto Start (needs 25 seconds without pressure).