![Datex-Ohmeda S/5 M-NIBP Technical Reference Manual Slot Download Page 19](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-nibp/s-5-m-nibp_technical-reference-manual-slot_513844019.webp)
NIBP Module, M-NIBP
Document No. 800 1015-1
15
17.
Keep the cuff on your arm and perform one NIBP measurement. Check that the module
gives a reasonable measuring result.
?
18.
Connect an infant cuff to cuff connector and wrap it around your fingers.
Start NIBP measurement and check that the module identifies the cuff, i.e. the text ‘Infant’
appears into the NIBP digit field. Cancel the measurement after the cuff identification.
?
19.
Perform electrical safety check and leakage current test.
?
20.
Check that the module functions normally after the performed electrical safety check.
?
21.
Clean the module with suitable detergent.
?
•
Fill in all necessary documents.