Datex-Ohmeda S/5 monitors
Document No. 800 1015-1
12
7.
Enter the NIBP module service menu:
Parameters - NIBP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check also that the NIBP board memories have passed the
internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
?
8.
Check the front panel LEDs and membrane keys.
Select
Buttons/Leds
Highlight the text
Auto ON
. Check that the LED for the autocycle measurement is turning on
and off on the module front panel when pressing the ComWheel. Check also the other LEDs
by selecting
Manual ON
,
STAT ON
and
Measur. ON
.
Press each of the module’s membrane keys at least for one second. Check that the pressed
key is identified, i.e. the text OFF changes to ON for the key in the menu.
?
9.
Check the pump and valves.
Select
Pneumatics
from the NIBP menu. Connect a pressure manometer to the NIBP
module cuff connector.
Highlight
Start Pump
and press the ComWheel. Check that the pump turns on and the
pressure inside the tubing system starts to increase. Stop the pump by pressing the
ComWheel again when the pressure reaches 280 mmHg.
Highlight
Open Exh1
. Press the ComWeel and check that the pressure inside the tubing
system starts to drop then press the ComWheel again. Check the other exhaust valve by the
same way by selecting
Open Exh2
from the menu.
If necessary, turn the pump on again for a moment to increase the pressure inside the
tubing system.
Highlight
Open Zerovalve
. Press the ComWheel and check that the pressure B1
≅
0 mmHg.
Close the zero valve by pressing the ComWheel again and check that the pressure B1 is
equal with the pressure B2.
Highlight
Set Valve
. Press the ComWheel and set the value under the text ‘Pulse Valve’ to
number 150 by turning the ComWheel. Press the ComWheel again and check that the
pressure inside the tubing system starts to drop. Finish the test by selecting
Previous Menu
.
?
Summary of Contents for S/5 M-NIBP
Page 2: ......
Page 36: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 32 This page intentionally left blank ...
Page 44: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 40 ...
Page 45: ...APPENDIX A NIBP Module M NIBP Document No 8001015 1 41 APPENDIX A ...
Page 46: ...Datex Ohmeda S 5 monitors Document No 800 1015 1 42 ...