![Datex-Ohmeda S/5 M-NIBP Technical Reference Manual Slot Download Page 48](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-nibp/s-5-m-nibp_technical-reference-manual-slot_513844048.webp)
Datex-Ohmeda S/5 monitors
A-2(2)
Document No. 8001015-1
13.
Watchdog timer
Allowed range
Adult
120...140 s
Infant
60...70 s
14.
Safety valve
B1
B2
Allowed range
‘Max press’ ADULT
290...330 mmHg
‘2 s after stop’ ADULT
290...330 mmHg
‘Max press’ INFANT
154...165 mmHg
‘2 s after stop’ INFANT
154...165 mmHg
OK
N.A.
Fail
OK
N.A.
Fail
15.
Cuff related messages
16.
Adult cuff detection
17.
Test measurement
18.
Infant cuff detection
Notes
19.
Electrical safety check
20.
Functioning after
electrical safety
check
21.
Final cleaning
Notes