DT9502
40
20171205-V3
7.7. The temperature value belonging to the 4 mA value
(05
menu item)
Function
The apparatus regulates the loop current between 4 and 20
mA in proportion with the measured temperature.
In this menu item any °C values between -200.0 °C and
600.0 °C could be assigned to the 4 mA loop current.
[Factory default setting: -200.0 °C]
Attention! In addition to selecting the temperature value
assigned to 4 mA, also the temperature value to be
assigned to 20 mA must be selected (The next menu item).
Sequence of operations
1. Enter the menu with the supervisor code.
Read chapter
7.1. Writing the code in
to see how the code
should be written in. On the display you see this:
.
2. Keep stepping with the
▲
button as long as you see this
menu item:
.
3. Enter the menu item by using the
ENTER
button.
4. The number
is visible on the display. Digit 2 is
blinking. The blinking digit indicates which of the digits is
selected. The selection can be modified by using the
▲
button.
6. Using the
►
button you can go to the next digit.
7. By using the
▲
button, you can increase the value of the
blinking digit:
,
,
,
,
,
,
,
,
,
, etc.
8. If you have selected the last digit (the digit on the right
side), and you press the
►
button, a blinking colon will
appear on the left side:
. This colon indicates,
that the positive or negative sign is being selected, and you
can now change it by the
▲
button.
9. With the
▲
button you can switch the negative sign on or
off.
•
= 200.0 °C
•
= -200.0 °C
Summary of Contents for DT9502
Page 1: ...DT9502 Intrinsically Safe Temperature Meter Transmitter Operating Instructions ...
Page 64: ...DT9502 64 20171205 V3 10 2 Application example ...
Page 84: ...DT9502 84 20171205 V3 10 9 ATEX Certification ...
Page 85: ...DT9502 20171205 V3 85 ...
Page 86: ...DT9502 86 20171205 V3 ...
Page 87: ...DT9502 20171205 V3 87 ...
Page 88: ...DT9502 88 20171205 V3 ...
Page 89: ...DT9502 20171205 V3 89 ...
Page 90: ...DT9502 90 20171205 V3 ...
Page 91: ...DT9502 20171205 V3 91 ...
Page 92: ......