6-76
Removal and Replacement
Magnetic Stripe Registration Sensor (Without Clear Card)
Removal Procedure
1.
Unplug the printer.
2.
Remove the ribbon cartridge and cleaning roller.
3.
Remove the
4.
Remove the
5.
Remove the
Magnetic Stripe Rocker Assembly
6.
Push the sensor's cable connector through the opening in the chassis (from
outside to inside).
7.
Carefully push down on the sensor from above and remove it through the
bottom of the chassis. Do not use sharp tools.
Replacement Notes
•
Make sure the sensor is oriented with the orange wire (emitter) toward the
rear of the printer.
•
Be careful not to damage the wires when pushing the sensor into position.
•
Make sure the sensor is fully seated.
•
Using the Diagnostics in chapter 5, calibrate the magnetic stripe head as
described in
Start Sentinel Position
.