MODEL 2234 DIGITAL FLOW COMPUTER ___________________________
Table 2-5. Error Code Diagnostic Chart
(Continued)
Error Code
Possible Cause
Check
Solution
09
Pressure out
of range for
density
calculation
2. Incorrect or
malfunctioning
pressure probe being
used for computing
density calculation.
1. Check that pressure
is between 200 and
2099.9 PSIA, read
code 2.
1. Check transducer
calibration.
10
Station
volume totals
stepping rate
is greater
than 25
pulses per
second
1. Flow rate too high,
totalizing factor set
too low.
1. Check read code 17
(TK) for proper factor
value (paragraph
3.6.5).
1. Enter a corrected
totalizer factor per
paragraph 2.6.3.1.
11
Invalid ratio
of pipe I.D.
to orifice
diameter
1. Incorrect setting of
inside line diameter
and/or line orifice
diameter.
1. Check value of
inside pipe diameter
for individual lines,
read codes 241, 242,
243, 244 and 245
(paragraph 3.6.3).
2. Check value of line
orifice diameter for
individual lines, read
codes 251, 252, 253,
254 and 255
(paragraph 3.6.3).
1. Enter correct
values(s) for pipe I.D.
and/or pipe orifice
diameter.