Illustration 78: Place the PAH sensor on DryCal-1
Place the PAH sensor on the DryCal-1 and press OK.
Illustration 79: Enter DryCal-1 PAH concentration
Enter the DryCal-1 concentration and press OK.
Illustration 80: Calibrating
Wait for the scaling calibration to finish.
Illustration 81: Confirm PAH calibration
Press OK to confirm and save the PAH calibration. Press Cancel to dismiss the PAH calibration.
PAH Calibration
6-**
Place the PAH sensor
on DryCal-1
Press OK when done
PAH Calibration
6-**
Enter DryCal-1
concentration
000.10
µ
g/L
PAH Calibration
6-**
Calibrating…
PAH Calibration
6-**
Confirm calibration?
Cancel
Ok