User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 45
Illustration 68: Switching to calibration mode
Wait until the sensor has switched to calibration mode.
Illustration 69: Place the PAH sensor on DryCal-0
Remove the PAH sensor by loosening the two M8 nuts holding the sensor in place. Before placing the PAH sensor on
the DryCAL-0 read the reference PAH value printed on the side of the DryCAL-0 and remember the value or write it
down for later use. Clean the optical part of the sensor with clean water using lint free cloth and place
it on the
DryCAL-0.
Illustration 70: Enter DryCal-0 PAH concentration
Press OK to enter DryCal-0 PAH concentration.
Illustration 71: Enter DryCal-0 PAH concentration
Toggle the digits by moving the cursor up, down, left and right.
PAH Calibration
6-**
Switching to calibration
mode
PAH Calibration
6-**
Place the PAH sensor
on DryCal-0
Press OK when done
PAH Calibration
6-**
Enter DryCal-0
concentration
0.10
µ
g/L
PAH Calibration
6-**
Enter DryCal-0
concentration
000.10
µ
g/L