User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 43
Illustration 63: Confirmation of the calibration
Press OK to confirm the calibration. The values shown on the LCP is the measured turbidity value of the TTurbCAL and
should be equal to the TTurbCAL reference turbidity concentration ± 5 % + 0,5 FNU.
Illustration 64: Saving the calibration
Wait for the calibration to be saved.
Illustration 65: Calibration saved
The calibration has been saved. Press OK to return to the Calibration menu.
Turb. Calibration
6-**
Confirm calibration
Calibrated value:
698.7 FNU
Cancel
Ok
Turb. Calibration
6-**
Saving…
Turb. Calibration
6-**
Calibration saved
Press OK to return to the
Calibration menu