User Guide
MES 1003 WMS
Doc no. 100837-MAN – Rev. E1 – Rev. date 2022-10
© Danfoss IXA A/S | Page 42
Illustration 57: Enter refernce turbidity
Toggle the digits by moving the cursor up, down, left and right.
Pressing OK before reference value is entered will start the calibration process, so be sure to check that the reference
value is correct before pressing OK.
Illustration 58: Enter reference turbidity
Illustration 59: Enter reference turbidity
Illustration 60: Enter reference turbidity
Press OK when the reference turbidity is entered.
Illustration 61: Enter reference turbidity
Illustration 62: Calibrating the turbidity sensor
Wait for the calibration to finish.
Turb. Calibration
6-**
Enter reference turbidity
724.1 FNU
0000.1
FNU
Turb. Calibration
6-**
Enter reference turbidity
724.5 FNU
0
6
00.1 FNU
Turb. Calibration
6-**
Enter reference turbidity
724.5 FNU
06
9
.1 FNU
Turb. Calibration
6-**
Enter reference turbidity
724.5 FNU
069
8
.1 FNU
Turb. Calibration
6-**
Enter reference turbidity
724.5 FNU
0698.
7
FNU
Turb. Calibration
6-**
Calibrating…