63
Spyder2 4k User’s Manual
03-032-10195-04
DALSA
Error Codes
Cod
e
Description
Suggested Cause
number outside ROI.
tap number specified in
cag
command is
outside region of interest.
29
Unable to calibrate offset. Tap
number outside ROI.
Analog offset could not be calibrated
because tap number specified in
cao
command is outside region of interest.
Informal Message Codes
Code
Description
Suggested Cause
1
INFO: CRC check failure while
attempting to restore
calibration status
Pixel calibration status will be set to NOT
CALIBRATED (internal micro EEROM
failure)
2
INFO: CRC check failure while
attempting to restore pixel
coefficients
All pixel coefficients will be reset to zero
(external EEROM (I2C) failure)
4
INFO: Flash memory ID error
Reported from boot process, only if DSP
configuration fails (possible cause:
communication error with serial flash
memory )
8
INFO: DSP configuration file
missing or corrupt
Reported from boot process, only if DSP
configuration fails (possible cause)
16
INFO: Serial communication
failure while accessing
external ADC chip
Reported from boot process if
communication verification fails with
external ADC chip (camera voltage
measurement, SPI)
32
INFO: Calibration may be out-
of-specification (PRNU
coefficient clipped)
ccp
: several coefficients were clipped
during the
ccp
command process
64
INFO: Calibration may be out-
of-specification (FPN
coefficient clipped)
ccf
: several coefficients were clipped
during the
ccf
command process
128
INFO: Calibration may be out-
of-specification (DO+FPN >
511)
ccf
: at least one pixel -> digital
FPN coefficient > 511
256
INFO: Changing analog
settings of calibration mode
voids pixel calibration
sao
,
sg
in calibration mode after
calibration has been done
512
INFO: For better calibration
results, run FPN calibration
first
ccp
:
ccp
called before
ccf
1024
INFO: Coefficient may be
inaccurate. A/D clipping has
occured.
Insufficient analog offset or too much gain
for given light.
Monitoring Task Message Codes
Code
Description
Suggested Cause
1
WARNING: One or more
At least one of voltages is out of
Summary of Contents for Spyder2 S2-2x-04K40
Page 10: ...10 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 38: ...38 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 52: ...52 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 56: ...56 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 74: ...74 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 76: ...76 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...