36
Spyder2 4k User’s Manual
03-032-10195-04
DALSA
sequence, use the command
els 1
. To disable the end-of-line sequence, use the
command
els 0
.
Use the
sut
and
slt
commands to set threshold values between 0 and 255 for 8 bit data
modes, or 0 to 1023 for 10-bit data modes.
3.18 Monitoring the Camera
The camera is equipped with a red/green LED used to display the operational status of
the camera. The following table summarizes the operating states of the camera and the
corresponding LED states:
LED State
Priority Camera
Status
Condition
Blinking RED
1
Error
Fatal hardware failure
Steady RED
2
Warning
Monitoring task failure
Blinking GREEN
3
Progress
Lengthy operation in progress. While
camera is blinking green, do not power
down camera or memory may be corrupted.
Steady GREEN
4
OK
Healthy
Note:
When more than one condition is active, the LED indicates the condition with the
highest priority. Error and warning states are accompanied by corresponding messages
further describing the current camera status.
Monitoring Tasks
The camera enters a warning state when any of camera's continuously running
monitoring tasks detects a failure. Use the
wed
i i
command, where the first parameter
is the monitoring task and the second parameter is the enable or disable value, to display
the status of all the defined monitory tasks (if no parameter is passed) and/or to
enable/disable specific monitoring tasks.
Note: If you are performing defect detection down to a single pixel, enabling monitoring
tasks is not advisable because monitoring can occur during active video and could
potentially provide erroneous data.
The following table lists the monitoring tasks:
Monitoring
task number
Cod
e
Monitoring Task
Description
0
All
Enable/disable all monitoring tasks
1
1
Voltage Monitoring
Monitors all camera voltages
2
2
Temperature Monitoring
Monitors camera temperature
3
4
External SYNC presence
Monitors presence of external SYNC
4
8
External PRIN presence
Monitors presence of external PRIN
5
16
Gain Out Of Spec
Monitoring
Monitors current setting of analog
gain
6
32
Line Rate Below 1 KHz
Monitors current line rate
Summary of Contents for Spyder2 S2-2x-04K40
Page 10: ...10 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 38: ...38 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 52: ...52 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 56: ...56 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 74: ...74 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...
Page 76: ...76 Spyder2 4k User s Manual 03 032 10195 04 DALSA ...