86
Appendix B: ASCII Commands
Notes
Full Name
Set Baud Rate
Mnemonic
SBR
Argument(s)
Baud rate
9600
19200
57600
115200
230400
460800
Description
Set baud rate
Notes
Send command and then change speed of HyperTerminal.
Full Name
Direction
Mnemonic
SCD
Argument(s)
Direction
Forward
Reverse
External – controlled by CC3 signal
Description
Set sensor scan direction
Notes
Full Name
Exposure Mode
Mnemonic
SEM
Argument(s)
Mode
Internal
External
Description
Set exposure time mode
Notes
In internal mode the exposure time is controlled by the SET command
In external mode the sensor is exposed while CC1 signal is high
External mode is only available when the trigger mode is also external (STM 1)
SEM 1 overrides internally generated independent exposure times
Full Name
Exposure Time
Mnemonic
SET
Argument(s)
Exposure time
4 000 to 3000000 [ns]
Description
Set internal exposure time
Notes
Line time > ( Exposure time + 2,000 ns )
Full Name
Mirroring
Mnemonic
SMM
Argument(s)
Mode
Off
Image is flipped on the vertical axis
Description
Set mirroring mode
Notes
Full Name
Pixel Format
Mnemonic
SPF
Argument(s)
Selector
0:RGB8
2:RGB12