7.7 Use of Waste Tank
The waste liquid tank is used to collect defrosting and defrosting liquid, tissue thawing liquid,
rinsing liquid and waste. In order to avoid the harm of waste water to people and the
environment, 200 ml of 10% formalin solution or other disinfection should be pre-installed in
the barrel. The discharge of wastewater waste should be conformed to the relevant regulations
of the hospital. Pull the bucket plate up to remove the waste container and the connected water
pipe can be removed.
Precautions:
Often observe the liquid level and pour out the waste liquid tank in time.
Waste must be treated in a correct and efficient manner and must comply with relevant
wastewater treatment regulations.
8. FAQ and Handling
The table below is the most common list of questions that may arise when using the instrument.
The table lists possible causes and solutions to these problems.
Fault
Cause
Solution
Tissue blocks fall off
Inadequate freezing
embedding agent
Increase the amount of
embedding agent
The trimming feed is too large
Use the trimming mode of the
device to trim and set the
trimming thickness to
20~40um.
Section softening/melting
Temperature of specimen,
blade, anti-roll plate and blade
holder is not low enough
Set a lower temperature
The glass door is opened too
much during sectioning, and
the operator breathes hot air to
progress the knife holder area
Adjust the appropriate opening
angle of the glass door, and the
operator should pay attention
to adjusting the slice posture
Uneven section
Insufficient sample freezing
Set a lower temperature
Foreign matter attached to the
blade or anti-roll plate
Clean and remove foreign
objects
The cutting edge is damaged
Switch blade cutting edge
position or change blade