Dakewe(Shenzhen)Medical Equipment Co,. Ltd
29
V4.1
means to be invalid. After the selection is valid, the device automatically turns on the
disinfection function when the disinfection time reaches the set time.
3. Indication 3 is starting disinfection time. Click the time black box to display . Choosing the
hour and minute. Press the number on the bottom right
to increase and decrease
the time. Press the “Save” button
to save the settings and press “Cancel” button
to abandon the settings.
Precautions:
1: When the UV lamp is turned on, opening the glass door system will temporarily turn off the
UV lamp, then close the glass door again, and the UV lamp will automatically turn on;
2: During the timed start of disinfection, the disinfection can be turned off manually. If the
manual end time is not reached, the disinfection will be turned off.
3: If the timed automatic disinfection is also performed during the manual opening of the
disinfection, it will not stopped after 35 minutes, but will stopped at the end of the timed
automatic disinfection.
(After setting the time, press the "Save" button to save, otherwise the setting is invalid.)
5.3.4 Time reset in the function interface
Double-click the icon "Reset"
in the lower left corner of the function interface, and the
date and time settings of the entire function interface will be automatically restored to the factory
default settings. One-click setting is convenient and quick.
5. 4 Compressor Overheat Protection Function
This is the compressor protection function in the system by default, and there is no corresponding
display prompt; when the compressor continues to run for more than 4 hours, it will stop for 3
minutes and then turn on again (if the glass door is opened or the semiconductor refrigeration is on
at this time, Time will be postponed).