Dakewe(Shenzhen)Medical Equipment Co,. Ltd
16
V4.1
4.
First Installation
4.1
Installation Requirements
Room temperature is always b5°C and +35°C.
Relative humidity is up to 60% without condensation.
The instrument is for indoor use only.
There shall be enough space around the hand wheel for easy operation.
For smooth operation, make sure that there are no other instruments that can cause vibration
near the microtome.
When placing the instrument, it must be fixed with the feet so as to avoid instability of the
instrument to affect the quality of the sections.
Avoid vibration, direct sunlight, and sudden temperature changes.
The installation site must be well ventilated, and no fire source is allowed.
4.2
Unpacking Tools
1. Electric screwdriver 2. Adjustable wrench 3. Utility knife 4. Allen wrench
Always wear protective gloves for installation and removal to avoid cuts.
Before unpacking, check whether the fasteners on the packing container are complete, and
whether the appearance is damaged or deformed. If the above situation occurs, mark it on
the delivery documents and check whether the goods are damaged.
Figure 4 Tools for unpacking