Dakewe(Shenzhen)Medical Equipment Co,. Ltd
12
V4.1
Attention: Within 100mm of the machine, 300mm of the vent,no other items can be
placed. The too high room temperature and the humidity will affect the cooling
capacity of the cryostat.
2.4 Instrument Handling
The machine is a movable floor-type device, with 4 casters on the bottom, 2
directional casters on the rear side, 2 universal casters on the front side, and 2 foot
cups on the front side directional caster attachment. The foot cup can be adjusted by
rotating.
When the whole machine needs to be fixedly placed, raise the foot cup so that the
front side directional casters are suspended and the whole machine is in a fixed state.
When the whole machine needs to move, lower the foot cup so that the front side
directional caster touches the ground, and the machine can be pushed at this time.
After unloading the transport package, adjust the front side directional caster to the
ground position and move to the place where it is placed. After it is in place, adjust
the height of the foot cup, adjust the front side directional caster to the suspended
state, and stabilize the machine horizontally.
2.5 Precautions for Use of the Instrument
The machine must be allowed to stand for two hours after installation in the
designated location.
Before starting the equipment, please check the voltage and frequency of the input
power to ensure that it conforms to the specification requirements on the instrument
nameplate (refer to figure 1).
After confirming that the power supply has a grounding wire (three-hole socket),
connect the unit and the power socket with the supplied power cord, and turn on the
power switch on the right of the machine to enter the working state.
Keep the hand wheel in the locked position if the machine is shutdown.
After the machine is stopped, the glass door should be opened to allow the water
vapor to evaporate as soon as possible to keep the cooling chamber dry.
After the UV lamp is damaged, it needs to be replaced as it is the same brand, model
and specification. Pay attention to safe conditions when replacing.
The components of the electrical system, such as the control panel, electric control
box, etc., must be maintained by professionals, not other personnel.
The refrigeration system is mainly composed of a compressor, a condenser, a filter,
an evaporator, etc., wherein the condenser (with a fan) adsorbs dust after long-term
use, affecting the cooling effect, and needs to open the ventilation plate on the right
side of the fuselage, Brush the dust on the condenser fins or blow it with high
pressure compressed air. (If the air filter on the right side is equipped with a dustproof
net, it must be cleaned regularly with a brush to ensure that the mesh is unblocked,
and the mesh blockage will reduce the cooling effect of the equipment)
The cooling chamber should be cleaned frequently and kept clean, but especially pay
attention that not to be scratched by the blade.