Dakewe(Shenzhen)Medical Equipment Co,. Ltd
38
V4.1
7. Cleaning and Maintenance
7.1 Daily finishing work
Lock hand wheel.
Remove the blade from the blade holder and put it back into the blade box.
Clear slicing waste.
Close sliding glass door.
Turn off the interior light.
Do not turn off the main switch, do not stop the refrigeration system.
Cleaning up the blade holder and brush holder.
7.2 Long-term Suspension of Finishing Work
Turn off the machine's main power switch.
Open the sliding glass door to dry the freezer.
Remove all the cartridges from the freezer.
Lock hand wheel.
Remove the blade on the blade holder.
Use a brush to remove all slicing waste.
Remove the waste tray, the front brush holder and blade holder for cleaning and disinfection.
Precautions:
Before the instrument is restarted, it must be checked whether all parts of the sliced movement
in the freezer are completely dry.
7.3 Cleaning Instrument
Generally, alcohol or alcohol-containing conventional disinfectants can be used for cleaning.
It is strictly forbidden to use acetone, xylene or other corrosive solvents to clean and disinfect
7.3.1 Before each cleanup, perform the following steps first
Raise the specimen clamp to the top position, then lock the hand wheel.
Loosen the loader locking screw and unplug the Sample disk.
Remove the blade from the blade holder, then put it back to the cartridge.
Remove the blade holder to clean up.
Remove the specimen from the specimen clamp. Use a dry brush to remove the section
waste.