MNL-00008 [5] 

Proprietary and Confidential Page 

47

 of 

55

 

 
Maximum Number of Treatments = 29 
 

SKIN TYPES I & II, ALL SCHEDULES 

 

% of 

MED 

% INCREASE OVER PREVIOUS DOSE 

TREATMENT # 

9  10  11  12  13  14  15  16  17  18  19  20  21  22  23  24  25  26  27  28 

29 

PRIMARY DOSE (MILLIJOULES) 

70 

10  10  10  10  10  10  10  10  10  10  10  10  10  9 

 

 

% OF PRIMARY DOSE 

SECONDARY DOSE (MILLIJOULES) 

50 

50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50 

50 

 

SKIN TYPES III & IV, ALL SCHEDULES 

 

% of 

MED 

% INCREASE OVER PREVIOUS DOSE 

TREATMENT # 

9  10  11  12  13  14  15  16  17  18  19  20  21  22  23  24  25  26  27  28 

29 

PRIMARY DOSE (MILLIJOULES) 

70 

15  15  15  15  15  15  15  15  15  15  15  14  13  12  11  10  9 

 

 

% OF PRIMARY DOSE 

SECONDARY DOSE (MILLIJOULES) 

50 

50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50 

50 

 

SKIN TYPES V & VI, ALL SCHEDULES 

 

% of 

MED 

% INCREASE OVER PREVIOUS DOSE 

TREATMENT # 

9  10  11  12  13  14  15  16  17  18  19  20  21  22  23  24  25  26  27  28 

29 

PRIMARY DOSE (MILLIJOULES) 

70 

15  15  15  15  15  15  15  15  15  15  15  15  15  14  13  12  11  10  9 

 

 

% OF PRIMARY DOSE 

SECONDARY DOSE (MILLIJOULES) 

50 

50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50  50 

50 

 

23.14

 

HI Dose Narrow Band UVB Maintenance Phase Protocols:  

 

1/WEEK 
 
Maximum Number of Treatments = 4 
 

ALL SKIN TYPES:

  Dose should be held steady at last clearance phase dose. 

 

Summary of Contents for UV Series

Page 1: ...MNL 00008 5 Proprietary and Confidential Page 1 of 55 ClearLink Controlled Phototherapy Equipment Clinical Operation Manual...

Page 2: ...ipment Control Systems 16 16 1 Dosimeter Instructions 3 Series NeoLux UV Series 7 Series M Series and 4 Series 16 16 1 1 Setting up a Treatment 16 16 1 2 Special Functions of the Integrated Dosimeter...

Page 3: ...amp Removal and Replacement 29 20 0 M Series CX and M Series CT Information 29 20 1 Electrical Requirements 29 20 2 Unpacking and Assembly 30 20 3 Special Notes 31 20 4 How to Position the Patient 31...

Page 4: ...011 3 week 2 week 46 23 11 LO Dose Narrow Band UVB Maintenance Phase Protocols 1 WEEK 46 23 12 LO Dose Narrow Band UVB Treatment Limits for Maintenance As Required Protocols 46 23 13 HI Dose Narrow Ba...

Page 5: ...our Customer Service department and your input will be channeled to the appropriate person The purpose of this manual is to instruct users on the proper methods of operation and general maintenance In...

Page 6: ...garis WARNING Do not use these devices for anything other than their intended purposes CAUTION Federal law restricts this device to sale by or on the order of a physician or any other practitioner lic...

Page 7: ...been tested and found to comply with the EMC limits of the international standard IEC 60601 1 2 These limits are designed to provide reasonable protection against interference in a typical medical ins...

Page 8: ...ry default key code needed to unlock the unit Figure 1 Unlock Code A warning label is affixed to your device in a prominent and easily readable position Please read the label carefully as it contains...

Page 9: ...finitions Table 1 Symbols Symbol Description DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION TYPE B EQUIPMENT 3 Series NeoLux M Series FAN 3 Series NeoLux Only LIGHTS...

Page 10: ...e delivery receipt that you will be asked to sign by the delivery driver Be sure to open the crate and verify that there is no damage to your unit before the driver leaves If it is not possible for yo...

Page 11: ...lamp Table 2 Lamp Specification Guide Applicable Units UVA UVA 1 Narrowband UVB Broadband UVB Blue Visible 3 Series NeoLux F72T12 BL HO RDC TL100W 10R TL 01 100W FS72 FS72T12 NB UVB HO RD C 100W FS72...

Page 12: ...cian nurse or any other practitioner licensed by the law of the state or country in which he she practices Training and experience in dermatology is important to provide holistic patient care This kno...

Page 13: ...ose This device is only to be used by authorized users Do not operate this device with a damaged cord or plug If appliance coupler or mains plug is used as the mains disconnect do not position the ME...

Page 14: ...xposed to flowing or splashing liquids of any kind If the device is inadvertently exposed to liquid it must be tested for safe function before being placed in operation again The device contains glass...

Page 15: ...ble to assist you Monday through Thursday 7 30am 8 00pm EST and Friday 8 30am 5 00pm EST 15 0 Checking the Power Output of Device The power output value of your device as measured at the factory is sh...

Page 16: ...ffective way to ensure the precision and consistency of each treatment 16 1 1 Setting up a Treatment 1 After unlocking device tap the Treat Patient button 2 Touch the arrows to set desired dose in mil...

Page 17: ...h variations and then integrates them into the dose treatment time This is a very effective way to ensure the precision of each treatment 16 2 1 Setting Up a Treatment 1 After unlocking device tap the...

Page 18: ...ls shown on the Power Output Certificate to determine estimated time equivalents needed to deliver doses in millijoules UVB or Joules UVA or visible light To determine a time equivalent consult the la...

Page 19: ...er the instructions 5 Confirm Treatment Time If Treatment Time is correct tap Enter Arrow to begin treatment If changes need to be made to Treatment Time tap Return Arrow to return to step 2 6 Followi...

Page 20: ...put is 4 0 mW cm2 and the dose to be delivered is 280 mJ then the appropriate treatment time will be 01 10 A partial time chart illustrating this is shown in mW cm2 mJoules 2 5 3 0 3 5 4 0 4 5 5 0 5 5...

Page 21: ...ll display progress To repeat the same dose for another exposure touch the Repeat button Note The system beeps to signal the end of the treatment and displays the delivered dose with the elapsed time...

Page 22: ...on is indicated with the Light icon and is used to start and pause stop the treatment The Blue fan button is indicated with a fan icon and is used to turn the fan on or off as desired 17 4 3 Series Ne...

Page 23: ...ecessary to safely support operation Table 5 Electrical Requirements and Figure 8 NEMA Outlets on page 23 define these requirements for each model variation in the UV Series Follow these specification...

Page 24: ...arrives Daavlin specifies that the junction box must be equipped with a safety disconnect switch This enclosure should be mounted to the wall against which the unit will stand at a height of four fee...

Page 25: ...m inside the machine once it has been moved to its final location 18 5 How to Position the Patient Enter the unit and close the doors In order to receive even and effective treatments it is important...

Page 26: ...her voltages may be shipped with no wall plug which would then have to be locally obtained and require an appropriate outlet that must be installed to meet local standards and requirements All power c...

Page 27: ...um of two lift the device including the foam shock absorbers from the box crate and lay it aside Do not remove the foam shock absorbers Note Your 7 series can weigh as much as 135 pounds Always use tw...

Page 28: ...permanent position 10 Install the power cord by inserting the male D shaped plug into the D shaped receptacle at the back of the device Then plug the unit into any grounded three prong electrical rece...

Page 29: ...until it clears the top socket then remove the lamp Reverse the process to install lamps and grid Note Please see the Lamp Inspection section on page 11 for the correct specifications of replacement...

Page 30: ...es CT Phototherapy equipment contain no operator replaceable parts NUMBER OF LAMPS VOLTS HERTZ AMP RATING FUSE OUTLET 10 110 120 60 6 3 3 Prong Grounded 10 220 240 50 60 3 15 Use Local Outlet FUSE TYP...

Page 31: ...If using the UVB lamps position your hands or feet in the back half of the unit 20 4 How to Position the Patient The hands feet or elbows should be placed through the arched exposure ports in the fron...

Page 32: ...steps 5 8 21 0 4 Series Information 21 1 Electrical Requirements Daavlin offers the 4 Series Phototherapy equipment with either 10 or 20 lamps Units can be equipped with UVA UVA 1 Blue Light Narrowba...

Page 33: ...near the ground and that lip of the foot is facing away from the unit For a picture of the proper foot alignment please see the cover of this manual 5 To secure the foot simply reinsert the four thum...

Page 34: ...m its corresponding lamp socket With one hand firmly press and hold the red button down While holding the red button down use your other hand to carefully lift the lamp up and out of the socket 5 With...

Page 35: ...nds on each side of the device and pull the 1 Series including the foam packers on each end out of the box Remove the foam packers Note Be sure to remove all of the accessories such as the manual cali...

Page 36: ...top of the clasp and use your index fingers to pull the sides of the clasp outward until it releases and the clasp can be removed 3 On top of each lamp socket there will be a red button that when pus...

Page 37: ...r 1 Series will have two outputs noted in the Current Value box of your certificate The higher of the two output values corresponds to the 0 in setting and the lower of the two values corresponds to t...

Page 38: ...measured from 9 inches 22cm away Note For hand and foot treatments or any other treatment that will be administered directly on the acrylic surface always use the higher output value to calculate your...

Page 39: ...0 5 0 5 0 UVA Skin Type I 3 week TREATMENT 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 PRIMARY DOSE JOULES 1 5 1 5 1 5 2 0 2 0 2 0 2 5 2 5 2 5 3 0 3 0 3 0 3 5 3 5 3 5 4 0 4 0 4...

Page 40: ...NDARY DOSE JOULES 1 0 1 5 2 0 2 5 3 0 3 5 4 0 4 5 5 0 5 5 6 0 6 5 7 0 7 5 8 0 8 0 8 0 8 0 8 0 8 0 8 0 8 0 8 0 8 0 8 0 UVA Skin Type II 2 week TREATMENT 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19...

Page 41: ...0 7 5 8 0 8 5 9 0 9 5 10 0 10 5 11 0 11 5 12 0 12 0 12 0 12 0 12 0 12 0 12 0 12 0 23 2 UVA Clearing Phase Protocols Skin Types IV VI Maximum Number of Treatments 25 UVA Skin Type IV 11011 TREATMENT 1...

Page 42: ...ULES 3 0 4 0 5 0 6 0 7 0 8 0 9 0 10 0 11 0 12 0 13 0 14 0 15 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 16 0 UVA Skin Type V 3 week TREATMENT 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18...

Page 43: ...13 5 14 5 15 5 16 5 17 5 18 5 19 5 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 20 0 SECONDARY DOSE JOULES 4 0 5 0 6 0 7 0 8 0 9 0 10 0 11 0 12 0 13 0 14 0 15 0 16 0 17 0 18 0 19 0 20 0 20 0 20...

Page 44: ...29 ALL SKIN TYPES ALL SCHEDULES of MED INCREASE OVER PREVIOUS DOSE TREATMENT 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 PRIMARY DOSE MILLIJOULES 30 50 40 30 20 19 1...

Page 45: ...25 26 27 28 29 PRIMARY DOSE MILLIJOULES 70 17 17 17 17 17 17 17 17 17 17 17 17 17 17 10 10 10 10 10 10 10 10 10 10 10 10 10 10 OF PRIMARY DOSE SECONDARY DOSE MILLIJOULES 50 50 50 50 50 50 50 50 50 50...

Page 46: ...ES 30 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 0 0 0 0 0 0 0 OF PRIMARY DOSE SECONDARY DOSE MILLIJOULES 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50 50...

Page 47: ...3 24 25 26 27 28 29 PRIMARY DOSE MILLIJOULES 70 15 15 15 15 15 15 15 15 15 15 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0 0 0 OF PRIMARY DOSE SECONDARY DOSE MILLIJOULES 50 50 50 50 50 50 50 50 50 50 50 50 5...

Page 48: ...nt Limits for Maintenance As Required Protocols Skin Type I II III IV V VI MED RANGE 50 300 mJ 100 400 mJ 150 500 mJ 200 600 mJ 250 700 mJ 300 800 mJ DOSE LIMITS with WARNING 1100 mJ 2000 mJ 4400 mJ 6...

Page 49: ...Cleaning Disinfection 24 2 1 General Cleaning For general use use a clean non abrasive cloth not paper towels and a mild cleaning solution such as Dawn Liquid Dishwashing Soap to gently wipe down the...

Page 50: ...ion please contact the Daavlin service department at 1 800 322 8546 for step by step instructions 25 0 EMC Guidelines Table 7 Electromagnetic Emissions Emissions Test Compliance EMC Environment Guide...

Page 51: ...553 MHz to 13 567 MHz 6 0 Vrms 26 057 MHz to 27 283 MHz 6 0 Vrms 40 66 MHz to 40 70 MHz 6 0 Vrms 0 15 MHz to 80 MHz 3 0 Vrms 6 765 MHz to 6 795 MHz 6 0 Vrms 13 553 MHz to 13 567 MHz 6 0 Vrms 26 057 MH...

Page 52: ...e survey should be less than the compliance level in each frequency range a Interference may occur in the vicinity of equipment marked with the following symbol Radiated RF IEC 61000 4 3 3 V m 80 MHz...

Page 53: ...applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80MHz and 800MHz the separat...

Page 54: ...nt parts on request 26 4 Warranty Service During the warranty period Daavlin will at Daavlin s option repair or replace any Equipment or components that appear to have been defective in material or wo...

Page 55: ...ories 26 6 Other Services Extended warranties are available and may be purchased from Daavlin s aftermarket sales department In the event that this Limited Warranty conflicts with other warranties inc...

Reviews: