Operating Specifications
Ambient Temperature:
15°C to 30°C (59°F to 86°F)
Relative Humidity:
10% to 95%, Non-condensing
Liquid Ingress Rating:
IPX0 (This device does not have protection
against ingress of water.)
Ocular Hazard Distance:
3 Meters (9.84 Feet)
Ambient Luminance:
250
–
500 lux
Atmospheric Pressure
70 kPa to 106 kPa
WARNING:
Equipment not suitable for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
5.0
Transport and Storage Specifications
Ambient
Temperature:
-40°C to 70°C (-40°F to 158°F)
Relative Humidity: 10% to 95%, Non-condensing
Atmospheric
Pressure:
50 kPa to 106 kPa
6.0
EMC Precautions
The devices contained in this manual have been tested and found to comply with the EMC limits
of the international standard IEC 60601-1-2. These limits are designed to provide reasonable
protection against interference in a typical medical installation. The system can radiate radio
frequency energy if not installed in accordance with the instructions and may cause harmful
interference to other devices in the vicinity. There is no guarantee that interference will not
occur in a particular installation. If the system does cause interference with other devices, which
can be determined by turning the system off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
•
Re-orient or relocate the receiving device
•
Increase the separation between the equipment
•
Connect the System into an outlet on a circuit different from that to which the other
device(s) is connected
•
Consult the manufacturer or field service technician for help