MNL-00008 [5]
Proprietary and Confidential
Page
17
of
55
•
If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device provides option to resume
interrupted treatment.
16.1.4
ClearLink Controller Range and Accuracy
Backup Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Dose Range:
UVA: 0.1 Joules - 999.9 Joules
UVB: 1 Millijoule - 9999 Millijoules
Accuracy:
+ / - 5%
Calibration:
1 year or every 100 hours of use
16.2
Dosimeter Instructions: 1 Series
Your Daavlin 1 Series features a built-in power meter that measures energy output. Once a dose
has been entered the unit will use the information gathered by the sensor to determine the exact
amount of time that each treatment will last. The dose is entered in millijoules for UVB and
Joules for UVA
The output of fluorescent lamps varies with age, use and other factors such as room temperature.
Integrated Dosimetry measures for such variations and then integrates them into the dose
treatment time. This is a very effective way to ensure the precision of each treatment.
16.2.1
Setting Up a Treatment
1.
After unlocking device, tap the “Treat Patient” button.
2.
Tap “0 inches” button or “9 inches” button depending on desired treatment distance.
3.
Touch the ▲/▼ arrows to set desired dose in millijoules (UVB) or Joules (UVA)
4.
Tap the Enter Arrow
to enter Treatment Confirmation screen.
5.
Ensure patient has put on goggles and positioned themselves per the instructions.
6.
Confirm Dose and Estimated Treatment Time.
•
If Dose and Estimated Treatment Time are correct, tap Enter Arrow
to begin
treatment.
•
If changes need to be made to Dose, tap Return Arrow
to return to step #3.
7.
Following delayed start, if enabled, treatment will begin, and screen will display
progress.
To repeat the same dose for another exposure, touch the Repeat button
Note:
The system beeps to signal the end of the treatment and displays the delivered
dose with the elapsed time.
16.2.2
Special Functions of the Integrated Dosimeter
•
To pause an active treatment, tap the Pause button
•
To resume a paused treatment, tap the Play button