5.3
Unpack and empty the column
Unpacking procedure
To unpack the Primer Support and empty Adjustable Oligo Column follow these steps:
Step
Action
1
Disconnect the column from the system.
2
Place a suitably dimensioned collecting vessel under the column to collect
the acetonitrile. Make sure to collect the acetonitrile directly in the vessel or
connect a piece of tubing at the column outlet to the vessel.
3
Apply a low pressure compressed air flow at the inlet of the column to allow
the acetonitrile to come out at the bottom of the column.
4
Continue the air flow until the Primer Support has reduced in volume and is
free from the column wall.
Note:
This will take approximately 30 minutes depending on the amount of Primer
Support in the column and the air flow.
5
Turn off the air flow.
6
Disconnect the top end piece by turning the adaptor counter clockwise until
it can be and lifted out.
7
Turn the column upside down and collect the Primer Support in a suitable
vessel.
8
Start the deprotection as described in the instruction manual for the
connected system.
5 Operation
5.3 Unpack and empty the column
24
Adjustable Oligo Column Operating Instructions 28967444 AC