R-net Omni2 Technical Manual | Chapter 8 – Diagnostics | 137
SK81935-01
3.25 RIGHT SIGNAL 1 SHORT CIRCUIT
This occurs when the Omni2 has detected that the Right signal from the Port 1 D-type connector has become short-circuit.
Refer to the Specifications chapter for connection details.
Check the operation of the SID and all cables and connectors between the Omni2 and the Port 1 D-type connector.
If the trip is still present after the above checks have been made, then the SID or Omni2 may be defective. Refer to Section 5.
3.26 LEFT SIGNAL 1 SHORT CIRCUIT
This occurs when the Omni2 has detected that the Left signal from the Port 1 D-type connector has become short-circuit.
Refer to the Specifications chapter for connection details.
Check the operation of the SID and all cables and connectors between the Omni2 and the Port 1 D-type connector.
If the trip is still present after the above checks have been made, then the SID or Omni2 may be defective. Refer to Section 5.
3.27 FORWARD SIGNAL 1 SHORT CIRCUIT
This occurs when the Omni2 has detected that the Forward signal from the Port 1 D-type connector has become short-circuit.
Refer to the Specifications chapter for connection details.
Check the operation of the SID and all cables and connectors between the Omni2 and the Port 1 D-type connector.
If the trip is still present after the above checks have been made, then the SID or Omni2 may be defective. Refer to Section 5.
3.28 REVERSE SIGNAL 1 SHORT CIRCUIT
This occurs when the Omni2 has detected that the Reverse signal from the Port 1 D-type connector has become short-circuit.
Refer to the Specifications chapter for connection details.
Check the operation of the SID and all cables and connectors between the Omni2 and the Port 1 D-type connector.
If the trip is still present after the above checks have been made, then the SID or Omni2 may be defective. Refer to Section 5.
3.29 FIFTH SWITCH 1 SHORT CIRCUIT
This occurs when the Omni2 has detected the Fifth Switch from the Port 1 D-type connector has become short-circuit. Refer
to the Specifications chapter for connection details.
Check the operation of the SID and all cables and connectors between the Omni2 and the Port 1 D-type connector.
If the trip is still present after the above checks have been made, then the SID or Omni2 may be defective. Refer to Section 5.