Measuring Calibration
To calibrate the pump integrated controller, user has to adjust two calibration points for any kind of parameter, i.e. pH or Redox
To enter in CAL menu, from measuring mode, press
OK button then the + button, on display CAL will
flash, press OK button again to confirm
On display will appear POINT 1
pH PROCEDURE
•
Dip the electrode in buffer solution 7 pH
•
Adjust the value on the display up to reach 7.00
with + and - buttons
•
Press OK to save the data. On display will
appear POINT 2
•
Dip the electrode after washing with tap water
in the second buffer solution; we suggest 4.00
or 9.00 pH.
•
Adjust the value on the display up to reach the
second buffer solution value with + and -
buttons.
•
Press OK to save the data
ALARMS setting
Overdosing alarm: For any kind of problems may occur in the plan (wrong calibration, dirty or broken probe, etc.) if the pump
doesn't reach the Setpoint in OVER ALARM TIME, the pump stops the dosing. On the display Alarm message will flash, Alarm Led
will flash.
To enter in ALARM menu follow the procedure
described in paragraph - PUMP SETUP (pag.
When ALARM menu is reached, press OK to enter
in submenus
80
ENGLISH
Summary of Contents for BOMBAPRO PH-RX
Page 2: ......
Page 7: ...Limpieza y mantenimiento del electrodo 30 DIBUJOS 117 Bomba 117 Cuerpo bomba 118 ESPAÑOL 7 ...
Page 32: ......
Page 35: ...DISEGNI 117 Pompa 117 Corpo Pompa 118 ITALIANO 35 ...
Page 60: ......
Page 63: ...DRAWINGS 117 Pump 117 Pump head 118 ENGLISH 63 ...
Page 88: ......
Page 91: ...DESSINS 117 Pompe 117 Corps de pompe 118 FRANÇAIS 91 ...
Page 116: ......
Page 119: ......
Page 120: ...CDMU 00203ML1A 03 2016 ...