Fig 54 - Po 1,95
Fig 55 - Po 2,05
5.4.3
FUNCTIONING CHECK WITH MEASURING KNOB ON 6
1
Position the functioning check accessory in position 6 (second
lateral marking) and the measuring knob on the corresponding
value 6.
2
Turn the measuring knob on the value + 5,95 (measure
equivalent to the width of the calibration sign on the
measuring knob).
3
The arm should be displaced in stop position facing the
examiner.
4
Turn the measuring knob on the value + 6,05 (measure
equivalent to the width of the calibration sign on the
measuring knob).
5
The arm should be displaced in stop position facing the patient.